EasyManua.ls Logo

Zeiss AT 030 - Validated Cleaning and Disinfecting Agents

Zeiss AT 030
88 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2 Safety notes Addendum
Instructions for use and disinfection of
tonometer measuring prisms
2.5 Validated cleaning and disinfecting agents
8 / 18 319300-9231-000-AddGA01-GB-240320
2.5 Validated cleaning and disinfecting agents
The following cleaning and disinfecting agents have been used in
several validations.
Cleaning agents
• Palmolive 0.08 %
Disinfectants
Product Concen-
tration
Exposure
time
Rinsing
Hydrogen
peroxide H
2
O
2
3 % 10 minutes Rinse with cold water
for 5 minutes
Sodium
hypochlorite
NaOCl
10 % 10 minutes Rinse with cold water
for 10 minutes

Table of Contents