2 Safety notes Addendum
Instructions for use and disinfection of
tonometer measuring prisms
2.5 Validated cleaning and disinfecting agents
8 / 18 319300-9231-000-AddGA01-GB-240320
2.5 Validated cleaning and disinfecting agents
The following cleaning and disinfecting agents have been used in
several validations.
Cleaning agents
• Palmolive 0.08 %
Disinfectants
Product Concen-
tration
Exposure
time
Rinsing
Hydrogen
peroxide H
2
O
2
3 % 10 minutes Rinse with cold water
for 5 minutes
Sodium
hypochlorite
NaOCl
10 % 10 minutes Rinse with cold water
for 10 minutes