General Information
Humphrey Field Analyzer 3 Service Manual 12 2660021151225 A
7. Remove the old expended lamp assembly. Insert the
replacement lamp into the housing. Note the notch in the
base of the lamp housing lines up with the pin to the right of
the assembly.
Note: Do NOT touch the glass part of the lamp with your
fingers, as this will shorten the life expectancy of the lamp. If
your finger touches the glass portion of the lamp, wipe the lamp
clean with a soft cloth.
Figure 6.
8. Slide plate back into position, tighten screw and replace
connector cable.
Figure 7.
9. To replace the top access cover, insert the cover into the
opening and press the cover down to latch it into place.