Error Codes
Display message What it means What to do
• The sample volume is too small
There may be a problem with the
test strip
• Meter error
Use a new test strip and try again with proper amount of
sample
Perform a control solution test using a new test strip. If the
test results are within the range printed on the test strip vial,
retest using a fresh blood sample and a new test strip
If the control solution result is out of range or the error
reappears, contact Zoetis Customer Care
• Meter error
Internal reference error
(data memory)
Firmware checksum error
(program memory)
Perform a control solution test using a new test strip. If the
test results are within the range printed on the test strip vial,
retest using blood and a new test strip
If the control solution result is out of range or the error
reappears, contact Zoetis Customer Care
Incorrect test procedure (e.g.,
applying blood on the test strip
before prompted by the meter)
There may be a problem with the
test strip
• Meter error
• The test strip is aected by moisture
Wait for the drop and strip icons to appear before applying
blood or control solution
Perform a control solution test using a new test strip. If the
test results are within the range printed on the test strip vial,
retest using blood and a new test strip
If the control solution result is out of range or the error
reappears, contact Zoetis Customer Care
There may be a problem with the
test strip
• Meter error
• Strip Sensing path broken
• Strip Type (OEM) error
Perform a control solution test using a new test strip. If the
test results are within the range printed on the test strip vial,
retest using blood and a new test strip
If the control solution result is out of range or the error
reappears, contact Zoetis Customer Care
Display message What it means What to do
There may be a problem with the test
strip connection with the meter
• Meter error
Perform a control solution test using a new test strip. If the
test results are within the range printed on the test strip vial,
retest using a fresh blood sample and a new test strip
If the control solution result is out of range or the error
reappears, contact Zoetis Customer Care
No calibration code Use the code strip attached to the test strip vial to set
the correct lot code.
Code Strip reading failed • Try to set the lot code again
If the code strip is illegible or broken,
contact Zoetis Customer Care
Operating temperature too low Try again after warming up the device.
Error Codes (continued)

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Zoetis AlphaTrack3 and is the answer not in the manual?

Zoetis AlphaTrack3 Specifications

General IconGeneral
BrandZoetis
ModelAlphaTrack3
CategoryBlood Glucose Meter
LanguageEnglish

Related product manuals