C
Cynthia PetersonSep 9, 2025
Why does my ZOLL aed pro defibrillator take more than 15 seconds to charge?
- HhunterhernandezSep 9, 2025
If your ZOLL Medical Equipment takes more than 15 seconds to charge, install a fully charged battery pack.
Why does my ZOLL aed pro defibrillator take more than 15 seconds to charge?
If your ZOLL Medical Equipment takes more than 15 seconds to charge, install a fully charged battery pack.
How to fix a "CHANGE BATTERY" message on ZOLL aed pro Medical Equipment?
If your ZOLL Medical Equipment displays the message "CHANGE BATTERY," replace the battery pack with a fully charged battery pack as soon as possible.
What to do if ZOLL Medical Equipment displays "PLUG IN DEFIB CABLE"?
If the ZOLL Medical Equipment displays the message "PLUG IN DEFIB CABLE", check the cable, and replace if necessary.
What to do if Ready indicator shows a red “X” on ZOLL Medical Equipment?
If your ZOLL Medical Equipment shows a red “X” while powered on, turn the unit off and then on again. Follow the prompts to resolve the problem. If the unit continues to fail, take it out of service, and contact ZOLL Technical Service.
What does "PLUG IN CABLE" mean on ZOLL aed pro and how to fix it?
If the ZOLL Medical Equipment displays the message "PLUG IN CABLE", ensure that the electrode cable is properly connected to the unit. Remove the cable and check for bent or broken pins. Replace the electrode cable if necessary.
How to troubleshoot the "CHECK ECG ELECTRODES ATTACH ECG ELECTRODES" message on ZOLL aed pro Medical Equipment?
If your ZOLL Medical Equipment shows the message "CHECK ECG ELECTRODES ATTACH ECG ELECTRODES", ensure that the ECG cable is connected to each electrode and to the unit. Also, make sure that the ECG electrodes are making good contact with the patient and are not dried out. Try replacing the ECG electrodes or the ECG cable.
Why does ECG data recording stop on my ZOLL aed pro Medical Equipment?
If ECG data recording stops on your ZOLL Medical Equipment, it may be because the unit was turned off and then turned on again within 10 seconds, which interrupts ECG recording.
What causes energy to not discharge when pressing the Shock button on ZOLL aed pro?
If the energy does not discharge when the Shock button is pressed on your ZOLL Medical Equipment, charge the defibrillator again, and deliver the shock while the charge-ready tone sounds. If the Shock button was pressed before the unit was fully charged, wait for the charge-ready tone and a flashing Shock button before pressing and holding the Shock button.
How to improve poor ECG signal level on ZOLL Medical Equipment?
If the ZOLL Medical Equipment shows a poor ECG signal level, replace the ECG electrodes and change their position on the patient.
What to do if ZOLL aed pro says "RELEASE LEFT SOFTKEY" or "RELEASE RIGHT SOFTKEY"?
If your ZOLL Medical Equipment displays the message "RELEASE LEFT SOFTKEY" or "RELEASE RIGHT SOFTKEY", release the softkey.
Device Type | Automated External Defibrillator (AED) |
---|---|
Display | LCD |
Operating Temperature | 0°C to 50°C (32°F to 122°F) |
ECG Display | Yes |
Waveform | Rectilinear Biphasic |
Voice Prompts | Yes |
Water Resistance | IP55 |
IP Rating | IP55 |
ECG Monitoring | Yes |
Warranty | 5 years |
Storage Temperature | -20°C to 60°C |
Energy Selection | Automatic, configurable |
CPR Feedback | Real CPR Help |
Data Recording | Yes |
Data Storage | Yes |
Connectivity | USB |
Shock Energy | Selectable 120J, 150J, 200J |
Energy Levels | Selectable 120J, 150J, 200J |