ZOLL Aed Plus - Service Manual

ZOLL Aed Plus
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Service Manual
9650-0308-01 Rev. A
March 2005

Other manuals for ZOLL Aed Plus

Questions and Answers

  • D
    Dennis MartinSep 17, 2025
    What to do if ZOLL Aed Plus shows '! UNIT FAILED'?
    • M
      Mark FullerSep 17, 2025
      If the unit displays '! UNIT FAILED', turn it off and then press and hold the ON/OFF button for at least 5 seconds to perform a self-test. Also, verify that the electrode cable is properly plugged into the patient cable connector. If the problem persists, replace the electrodes.
  • D
    Danny FloresAug 25, 2025
    How to fix 'CHANGE BATTERIES' prompt on ZOLL Aed Plus Medical Equipment?
    • C
      Colleen GibsonAug 25, 2025
      If your ZOLL Medical Equipment prompts you to 'CHANGE BATTERIES,' replace all batteries with new batteries at the same time. Remember to press the Battery Reset Button when prompted.
  • J
    Joshua KingAug 25, 2025
    How to fix the 'CHANGE BATTERIES' prompt on ZOLL Aed Plus Medical Equipment?
    • R
      Randy CareyAug 26, 2025
      To resolve the 'CHANGE BATTERIES' prompt on your ZOLL Medical Equipment, replace all batteries with new batteries at the same time. Then, press the Battery Reset Button when prompted.
  • A
    Angela SimpsonAug 26, 2025
    What to do if ZOLL Aed Plus Medical Equipment says 'ANALYSIS HALTED. KEEP VICTIM STILL'?
    • E
      Eric JohnsonAug 26, 2025
      If the ZOLL Medical Equipment displays 'ANALYSIS HALTED. KEEP VICTIM STILL', it means excessive artifact was detected during ECG analysis. Ensure the victim is motionless during ECG analysis, and that no one touches the victim during analysis. If you're using the device in an emergency vehicle, bring the vehicle to a halt before performing ECG analysis.
  • T
    thomascindySep 18, 2025
    Why was no shock delivered by my ZOLL Medical Equipment?
    • R
      Ronald JohnsonSep 18, 2025
      A shock might not be delivered for these reasons: 1. The Treatment button wasn't pressed within 30 seconds of being prompted. 2. An internal error occurred. If you failed to press the button in time, make sure to press it when prompted within 30 seconds. If an internal error occurred, remove the unit from service and contact Technical Service.
  • J
    juan94Sep 23, 2025
    What does '! RELEASE SHOCK BUTTON' mean on ZOLL Aed Plus?
    • W
      wmillerSep 23, 2025
      If you see '! RELEASE SHOCK BUTTON' (or '! RELEASE TREATMENT BUTTON'), release the Treatment button. Do not press the button until the charge-ready tone sounds and the button begins flashing.
  • M
    murphyrobinSep 9, 2025
    What to do if the ZOLL Aed Plus prompts 'PLUG IN CABLE'?
    • M
      Michael JohnsonSep 9, 2025
      If the ZOLL Medical Equipment prompts 'PLUG IN CABLE', check the cable connection between the electrodes and the unit.
  • P
    pgravesSep 10, 2025
    What to do if ZOLL Medical Equipment prompts 'RELEASE SHOCK BUTTON'?
    • C
      Cynthia ParsonsSep 10, 2025
      If the ZOLL Medical Equipment prompts 'RELEASE SHOCK BUTTON', release the Shock Button, then press and hold it again until discharge occurs.
  • A
    Andrew ClarkSep 3, 2025
    What to do if ZOLL Aed Plus prompts to 'RELEASE SHOCK BUTTON'?
    • M
      Martin ThompsonSep 3, 2025
      If your ZOLL Medical Equipment prompts you to 'RELEASE SHOCK BUTTON', release the Shock Button, then press and hold the Shock Button until discharge occurs. If the voice prompt continues, consider contacting ZOLL Technical Service.
  • J
    Jerry HillSep 1, 2025
    What to do if ZOLL Aed Plus self test failed?
    • M
      Mr. Gregory HintonSep 1, 2025
      If the self-test failed on your ZOLL Medical Equipment, initiate a manual self-test by pressing and holding the ON/OFF Button for more than 5 seconds. Attempt to repair the device by replacing the batteries or electrodes. If the AED Plus fails the test again, remove the AED Plus from service.

Summary

Preface

Safety Considerations

Provides warnings about servicing the unit and potential lethal shocks.

Technical Service

Contact information and procedures for obtaining technical service for the AED Plus unit.

Chapter 1 Maintenance Tests

Physical Inspection

Procedure to ensure the unit shows no signs of damage or excessive wear.

Preventive Maintenance Test

Verifies the AED Plus unit is working properly and ready for use.

Manual Self-Test

Initiates the unit's self-diagnostic procedure to verify key functions.

Chapter 2 Troubleshooting

Troubleshooting the AED Plus Unit

Provides common issues, solutions, and operator/technical actions for problems.

AED Plus Error Log Messages

Lists critical error log messages that indicate the unit requires technical service.

Appendix A Maintenance Tests Checklist

Instructions for Checklist Use

Guidance on using the maintenance checklist for record-keeping.

ZOLL Aed Plus Specifications

General IconGeneral
Device TypeAutomated External Defibrillator (AED)
WaveformRectilinear Biphasic
Electrode Shelf Life5 years
IP RatingIP55 (protected against dust and water jets)
Warranty5 years
Energy120J, 150J, 200J
CPR FeedbackReal CPR Help
Battery Type10 x 123A lithium batteries
Battery Life5 years
DisplayLCD
Weight6.7 lbs (3.1 kg)
Dimensions5.25 in (H) x 9.50 in (W) x 11.50 in (D) / 13.3 cm (H) x 24.1 cm (W) x 29.2 cm (D)
Energy SelectionFixed
Voice PromptsYes
Data RecordingInternal memory
Operating Temperature0°C to 50°C (32°F to 122°F)
Storage Temperature-4°F to 140°F (-20°C to 60°C)
Relative Humidity10% to 95% non-condensing
AltitudeUp to 15, 000 feet (4, 572 meters)

Related product manuals