Service Manual
9650-0308-01 Rev. A
March 2005

Other manuals for ZOLL Aed Plus

Questions and Answers

  • R
    Randy GordonSep 12, 2025
    Why is my ZOLL Medical Equipment beeping when off?
    • S
      Sandra JohnsonSep 12, 2025
      If your ZOLL Medical Equipment is beeping when it's turned off, remove it from service and replace the batteries. Make sure to replace all batteries at the same time. Press the battery reset button after replacing them. If the beeping continues after these steps, contact ZOLL Technical Service.
  • Q
    qariasSep 12, 2025
    Why does my ZOLL Aed Plus say 'ANALYSIS HALTED. KEEP PATIENT STILL'?
    • C
      Curtis NewmanSep 12, 2025
      If the ZOLL Medical Equipment displays the 'ANALYSIS HALTED. KEEP PATIENT STILL' prompt, it means excessive artifact was detected during ECG analysis. To resolve this, ensure the patient remains motionless during ECG analysis, and avoid touching the patient. If you're using the device in an emergency vehicle, stop the vehicle before performing ECG analyses.
  • P
    pgravesSep 10, 2025
    What to do if ZOLL Medical Equipment prompts 'RELEASE SHOCK BUTTON'?
    • C
      Cynthia ParsonsSep 10, 2025
      If the ZOLL Medical Equipment prompts 'RELEASE SHOCK BUTTON', release the Shock Button, then press and hold it again until discharge occurs.
  • M
    murphyrobinSep 9, 2025
    What to do if the ZOLL Aed Plus prompts 'PLUG IN CABLE'?
    • M
      Michael JohnsonSep 9, 2025
      If the ZOLL Medical Equipment prompts 'PLUG IN CABLE', check the cable connection between the electrodes and the unit.
  • T
    Thomas RileySep 8, 2025
    What to do if my ZOLL Aed Plus Medical Equipment is beeping when off?
    • H
      Heather BeckerSep 8, 2025
      If your ZOLL Medical Equipment is beeping when turned off, remove the unit from service and replace the batteries. Ensure you replace all batteries at the same time with new ones and press the Battery Reset Button when prompted. If the beeping continues, further assistance is needed.
  • J
    jennifer27Aug 23, 2025
    How to change batteries in ZOLL Aed Plus?
    • E
      Elizabeth DuncanAug 23, 2025
      If your ZOLL Medical Equipment prompts 'CHANGE BATTERIES', replace all batteries with new ones at the same time. Remember to press the Battery Reset Button when prompted after the replacement.
  • M
    Michelle BrownSep 6, 2025
    What to do if the ZOLL Aed Plus Medical Equipment displays 'ANALYSIS HALTED. KEEP PATIENT STILL'?
    • S
      Shannon FergusonSep 6, 2025
      If the ZOLL Medical Equipment displays 'ANALYSIS HALTED. KEEP PATIENT STILL', it means excessive artifact was detected during ECG analysis. Ensure the patient is motionless during ECG analysis and that no one touches the patient during analysis. If you're using the device in an emergency vehicle, stop the vehicle before performing ECG analysis. Keep the patient still.
  • V
    Valerie AllenSep 5, 2025
    What to do if ZOLL Aed Plus shows a red 'X' or beeps when off?
    • M
      murphyrobinSep 5, 2025
      If your ZOLL Medical Equipment is showing a red “X” in the status indicator window or making a beeping noise when the unit is off, you should: 1. Perform a manual test. 2. Check that the cable is properly attached to the unit. 3. Replace the batteries. If the unit still does not operate correctly after these steps, remove it from service.
  • A
    adriennefisherSep 4, 2025
    What to do if ZOLL Medical Equipment shows a red 'X' in the status indicator window when on?
    • H
      Hannah McdonaldSep 4, 2025
      If the ZOLL Medical Equipment shows a red “X” in the status indicator window when the unit is on, power cycle the unit. If the red “X” is still present, remove the unit from service.
  • A
    Andrew ClarkSep 3, 2025
    What to do if ZOLL Aed Plus prompts to 'RELEASE SHOCK BUTTON'?
    • M
      Martin ThompsonSep 3, 2025
      If your ZOLL Medical Equipment prompts you to 'RELEASE SHOCK BUTTON', release the Shock Button, then press and hold the Shock Button until discharge occurs. If the voice prompt continues, consider contacting ZOLL Technical Service.

Summary

Preface

Safety Considerations

Provides warnings about servicing the unit and potential lethal shocks.

Technical Service

Contact information and procedures for obtaining technical service for the AED Plus unit.

Chapter 1 Maintenance Tests

Physical Inspection

Procedure to ensure the unit shows no signs of damage or excessive wear.

Preventive Maintenance Test

Verifies the AED Plus unit is working properly and ready for use.

Manual Self-Test

Initiates the unit's self-diagnostic procedure to verify key functions.

Chapter 2 Troubleshooting

Troubleshooting the AED Plus Unit

Provides common issues, solutions, and operator/technical actions for problems.

AED Plus Error Log Messages

Lists critical error log messages that indicate the unit requires technical service.

Appendix A Maintenance Tests Checklist

Instructions for Checklist Use

Guidance on using the maintenance checklist for record-keeping.

ZOLL Aed Plus Specifications

General IconGeneral
Device TypeAutomated External Defibrillator (AED)
WaveformRectilinear Biphasic
Electrode Shelf Life5 years
IP RatingIP55 (protected against dust and water jets)
Warranty5 years
Energy120J, 150J, 200J
CPR FeedbackReal CPR Help
Battery Type10 x 123A lithium batteries
Battery Life5 years
DisplayLCD
Weight6.7 lbs (3.1 kg)
Dimensions5.25 in (H) x 9.50 in (W) x 11.50 in (D) / 13.3 cm (H) x 24.1 cm (W) x 29.2 cm (D)
Energy SelectionFixed
Voice PromptsYes
Data RecordingInternal memory
Operating Temperature0°C to 50°C (32°F to 122°F)
Storage Temperature-4°F to 140°F (-20°C to 60°C)
Relative Humidity10% to 95% non-condensing
AltitudeUp to 15, 000 feet (4, 572 meters)