ZOLL Aed Plus - Service Manual

ZOLL Aed Plus
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Service Manual
9650-0308-01 Rev. A
March 2005

Other manuals for ZOLL Aed Plus

Questions and Answers

  • N
    Nicole SwansonSep 2, 2025
    How to resolve 'PLUG IN CABLE' prompt on ZOLL Medical Equipment?
    • J
      Joshua AcostaSep 2, 2025
      If your ZOLL Medical Equipment displays a 'PLUG IN CABLE' prompt, check the cable connection between the electrodes and the AED Plus.
  • J
    Jerry HillSep 1, 2025
    What to do if ZOLL Aed Plus self test failed?
    • M
      Mr. Gregory HintonSep 1, 2025
      If the self-test failed on your ZOLL Medical Equipment, initiate a manual self-test by pressing and holding the ON/OFF Button for more than 5 seconds. Attempt to repair the device by replacing the batteries or electrodes. If the AED Plus fails the test again, remove the AED Plus from service.
  • K
    Kimberly EricksonAug 31, 2025
    What to do if there is a red “X” in the status indicator window on ZOLL Aed Plus Medical Equipment?
    • M
      Mrs. Maria MooneyAug 31, 2025
      If you see a red “X” in the status indicator window of your ZOLL Medical Equipment, initiate a manual self-test by pressing and holding the ON/OFF Button for more than 5 seconds. Check if the cable is properly attached to the AED Plus, or replace the electrodes. Cycle the power on the AED Plus by turning the unit off and then on again. Replace all batteries at the same time with new batteries that are less than 1 year old and press the Battery Reset Button when prompted. If the AED Plus still does not operate correctly, remove the unit from service.
  • Z
    zriveraAug 30, 2025
    Why does my ZOLL Aed Plus say 'ANALYSIS HALTED. KEEP VICTIM STILL'?
    • B
      burgessnathanAug 30, 2025
      The 'ANALYSIS HALTED. KEEP VICTIM STILL' prompt on your ZOLL Medical Equipment indicates that excessive artifact was detected during ECG analysis. Ensure the victim is motionless during ECG analysis and that no one is touching the victim. If using the AED Plus in an emergency vehicle, bring the vehicle to a halt before performing ECG analysis. The victim must be still.
  • P
    pmartinezAug 29, 2025
    How to fix a failed self test on ZOLL Aed Plus?
    • P
      Pamela RobinsonAug 29, 2025
      If the self-test fails on your ZOLL Medical Equipment, initiate a manual self-test by pressing and holding the ON/OFF Button for more than 5 seconds. Then attempt to repair the device by replacing the batteries or electrodes. If the device fails the test again, remove it from service and contact ZOLL Technical Service.
  • C
    Corey JonesAug 28, 2025
    What to do if ZOLL Aed Plus Medical Equipment says 'PLUG IN CABLE'?
    • W
      wpatrickAug 28, 2025
      If your ZOLL Medical Equipment prompts 'PLUG IN CABLE', check the cable connection between the electrodes and the device.
  • T
    Terri JensenAug 27, 2025
    What to do if ZOLL Aed Plus Medical Equipment has a red 'X'?
    • D
      Dawn SmithAug 27, 2025
      If your ZOLL Medical Equipment displays a 'Red “X”' in the status indicator window, initiate a manual self-test by pressing and holding the ON/OFF Button for more than 5 seconds. Also, check if the cable is properly attached or replace the electrodes. You can also try cycling the power by turning the unit off and then on again. Replace all batteries with new ones that are less than 1 year old and press the Battery Reset Button when prompted. If the issue persists, remove the unit from service and contact ZOLL Technical Service.
  • A
    Angela SimpsonAug 26, 2025
    What to do if ZOLL Aed Plus Medical Equipment says 'ANALYSIS HALTED. KEEP VICTIM STILL'?
    • E
      Eric JohnsonAug 26, 2025
      If the ZOLL Medical Equipment displays 'ANALYSIS HALTED. KEEP VICTIM STILL', it means excessive artifact was detected during ECG analysis. Ensure the victim is motionless during ECG analysis, and that no one touches the victim during analysis. If you're using the device in an emergency vehicle, bring the vehicle to a halt before performing ECG analysis.
  • J
    Joshua KingAug 25, 2025
    How to fix the 'CHANGE BATTERIES' prompt on ZOLL Aed Plus Medical Equipment?
    • R
      Randy CareyAug 26, 2025
      To resolve the 'CHANGE BATTERIES' prompt on your ZOLL Medical Equipment, replace all batteries with new batteries at the same time. Then, press the Battery Reset Button when prompted.
  • D
    Danny FloresAug 25, 2025
    How to fix 'CHANGE BATTERIES' prompt on ZOLL Aed Plus Medical Equipment?
    • C
      Colleen GibsonAug 25, 2025
      If your ZOLL Medical Equipment prompts you to 'CHANGE BATTERIES,' replace all batteries with new batteries at the same time. Remember to press the Battery Reset Button when prompted.

Summary

Preface

Safety Considerations

Provides warnings about servicing the unit and potential lethal shocks.

Technical Service

Contact information and procedures for obtaining technical service for the AED Plus unit.

Chapter 1 Maintenance Tests

Physical Inspection

Procedure to ensure the unit shows no signs of damage or excessive wear.

Preventive Maintenance Test

Verifies the AED Plus unit is working properly and ready for use.

Manual Self-Test

Initiates the unit's self-diagnostic procedure to verify key functions.

Chapter 2 Troubleshooting

Troubleshooting the AED Plus Unit

Provides common issues, solutions, and operator/technical actions for problems.

AED Plus Error Log Messages

Lists critical error log messages that indicate the unit requires technical service.

Appendix A Maintenance Tests Checklist

Instructions for Checklist Use

Guidance on using the maintenance checklist for record-keeping.

ZOLL Aed Plus Specifications

General IconGeneral
Device TypeAutomated External Defibrillator (AED)
WaveformRectilinear Biphasic
Electrode Shelf Life5 years
IP RatingIP55 (protected against dust and water jets)
Warranty5 years
Energy120J, 150J, 200J
CPR FeedbackReal CPR Help
Battery Type10 x 123A lithium batteries
Battery Life5 years
DisplayLCD
Weight6.7 lbs (3.1 kg)
Dimensions5.25 in (H) x 9.50 in (W) x 11.50 in (D) / 13.3 cm (H) x 24.1 cm (W) x 29.2 cm (D)
Energy SelectionFixed
Voice PromptsYes
Data RecordingInternal memory
Operating Temperature0°C to 50°C (32°F to 122°F)
Storage Temperature-4°F to 140°F (-20°C to 60°C)
Relative Humidity10% to 95% non-condensing
AltitudeUp to 15, 000 feet (4, 572 meters)

Related product manuals