Title: Defibrillator – Zoll E and M Date: February 8, 2021
By: Zoll Chelmsford, MA USA
File = Zoll Guideline
DISCLAIMER: THIS PROCEDURE PROVIDED "AS IS" AND WITH
POSSIBLE FAULTS. USER MUST VERIFY BEFORE USE. NEITHER
PROVIDER NOR WEBSITE ASSUMES ANY RESPONSIBILITY FOR ITS USE.
1. General
Service Guideline for Zoll Series M and E with ECG, SP02, NIBP.
2. Reference Documents
Operator Manual M Series 9650-0200-01
Service Manual M Series 9650-0450-01
Configuration Guide M Series 9650-0201-01
Operator Manual E 9650-1210-06
Service Manual E Series 9650-1450-01
Configuration Guide E Series 9650-1201-01
3. Accessories (see Appendix for photos)
“MFC” HV Paddle Cable, Hard Paddles or “Padz® Adult” See Appendix
Recorder ECG Paper (90 x 90 x 200 Z-fold)
ECG Cable: 3 lead, 5 lead (cable uses normal 6-pin connector)
SP02 Sensor/Cable (Massimo normally supplied)
CO2 Sensor/Cable, and data card – not addressed here.
NIBP Cuff/Tubing – standard issue from various sources.
4. Test Fixtures
ECG Simulator; Defibrillator Analyzer if the Dummy Load not available.
Note: Some units are manual mode only (no AED option); other units have both modes and can be configured to turn-on
either in manual mode, or AED mode. The following checks are written with the unit in manual mode, which is the
factory default. Other various user options are beyond the scope of this document. Warning: Performing a
“factory defaults”(a reset) to select manual mode at turn-on, will affect any custom configurations.
5. Basic PM Procedure (every 6 months per Zoll)
5.1. Physical InspectionNote paddles, case & power cord integrity, and battery age. Does handle works
properly? Recorder drawer opens and closes properly? Input connectors clean and undamaged? No
cuts on the accessory cables.
5.2. Turn On / Self Test
After connection to AC power, note the CHARGER ON indicator. Turn the defibrillator ‘ON’ by turning
the rotary knob on the main panel.
5.3. Battery Diagnostics / Info:
The E Series products use easily replaced sealed lead-acid or lithium-ion battery packs when new and
fully charged. They provide at least 2.5 hours of monitoring. M-Series use only sealed lead-acid.
The amber or green lights (below the screen) will illuminate after plugging the instrument to the mains.
If both lights flash ON/OFF, no battery is installed or the unit is defective. With a new battery, the green
LED usually lights within a few minutes.
Note: For a Zoll M or E Series compatible battery (10V/2.5Ah), one verified item is R&D Batteries, item #5603 –
sealed lead acid.
Zoll M
Zoll E

Other manuals for ZOLL M Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL M Series and is the answer not in the manual?

ZOLL M Series Specifications

General IconGeneral
Pacer Rate30 - 180 ppm
Display TypeLCD
Defibrillator TypeManual and AED
Monitor DisplayHigh-resolution LCD
ECG Monitoring3-lead
PacingDemand
Battery TypeLithium-ion
DisplayHigh-resolution LCD
WaveformBiphasic Truncated Exponential (BTE)
Monitoring ParametersECG, SpO2, EtCO2, NIBP, temperature
ConnectivityData port for data transfer
Pacer TypeTranscutaneous
Pacer Output0 - 140 mA
PrinterIntegrated thermal
Energy Selection2, 3, 5, 7, 10, 15, 20, 30, 50, 70, 100, 120, 150, 200 Joules