16
c) Position and mount both remote
input jacks in convenient locations.
Fig. 16 - Inserting Model 733 plug into rear of
Model 724
Section 11.
Verification Procedure for
the Model 724 Monitor
The Model 724 Monitor cannot be
recalibrated after the initial factory
calibration. However, the following
steps can be used to determine if the
Model 724 is operating within its
specifications.
Equipment Needed:
• Small blade screwdriver.
Resistance Substitution Box (RSB),
1 Ohm to 45 Megohms ±1%.
Fig. 17 - Model 724 wire connection diagram for two work stations.