The operator may select the memory location to store the depth readings
by pressing Mem Select [SK]. When the preferred location appears on the
screen, press Save [SK]. The screen and memory location will populate
with the current information.
Each memory location can be reviewed by pressing Mem Select [SK].
Press Locate [5] to return to Marker Locate Mode.
Depth of Passive, Non-iD Marker
Lower the tip of the receiver to the ground over the targeted marker.
Press Depth [SK].
The receiver will examine the targeted marker. (Calculating signal, please wait)
The screen will instruct the operator to raise the unit 6 inches (15.2 cm)
from the ground.
Note: This 6 inch rise must be exact for the depth reading to be accurate.
Raise the unit 6 inches. Press Depth [SK] key again. The estimated depth of
the marker from ground level will display on the screen.
Five depth readings can be saved with the time, and date.
To access the memory locations, press Mem Select [SK].