This is a quick guide, further details are in the 3M
Prevena
Therapy Clinician Guide
Alert type How to x
Leak Alert
One beep, one solid
yellow light
To ensure proper application of the 3M
Prevena
Incision Dressing, the Prevena
125 Therapy Unit
provides the Visicheck
Feature. Double press ON/OFF button.
- One light: Best
- Two lights: Good
- Three lights: Marginal
If the Visicheck
Feature indicates a “Marginal” leak rate condition, follow these steps:
- Ensure canister is locked into the unit
- Check the tubing connectors to make sure they are fully engaged and locked
- Use nger to smooth out any creases in the adhesive
- Use Prevena Incision Dressing Patch Strips (located in the dressing package) to help seal leaks.
- Slowly press rmly around dressing edges
Canister Full Alert
Two beeps, one solid
yellow light
Check uid level in the canister
1. If the canister isn’t full check for kinks or pinched tubing
2. If the canister is full then turn the unit o
3. Call 1-800-275-4524
Low Battery Alert
Low- One slow beep, one
solid yellow light
Critical- One beep,
repeating rapidly,
increasing volume, two
solid yellow lights
This therapy unit comes with three AA batteries
1. Hold ve second to turn o
2. Press and push the panel on the back of the unit to open
3. Insert three AA batteries (lithium batteries recommended)
4. Press and push to close panel
5. Hold two seconds to turn on
System Error Alert
One Beep, repeating
rapidly, increasing in
volume, two solid yellow
lights
1. Hold for ve seconds to turn o
2. Hold two seconds to turn back on
3. If the therapy is still not running call 1-800-275-4524
Device Lifecycle
Expired
Three lights show life
cycle is up. There will be
a repeated beep for 15
seconds and then the unit
will shut o.
Call 1-800-275-4524
Life cycle=192 hours (eight days)
3M
Prevena
125 Therapy Unit
indicators and alerts
Caution
Indicator
Audible Alarm
(muted)
Battery
Level
Quick Guide
Incision Therapy
Question and Answer IconNeed help?

Do you have a question about the 3M Prevena 125 and is the answer not in the manual?

Summary

3 M Prevena Therapy Unit Indicators and Alerts Guide

Leak Alert Troubleshooting

One beep, one solid yellow light; check Visicheck feature for proper application.

Canister Full Alert Fixes

Two beeps, one solid yellow light; check canister, tubing, or call for service.

Low Battery Alert Indicators

Beeps and yellow lights indicate low battery; replace AA batteries.

System Error Alert Resolution

Repeating beep, solid yellow lights indicate system error; troubleshoot or call.

Device Lifecycle Expired Notification

Three lights and beep signal end of life; unit shuts off.

Blockage Alert Detection

Audible beeps and yellow light indicate blockage; check canister, tubing, clamp.

Alert Mute Button Functionality

Press and hold mute button to silence audible alerts for two minutes.

Therapy Ended Status

Solid yellow LED, beeps indicate therapy completion; unit will turn off.

Summary

3 M Prevena Therapy Unit Indicators and Alerts Guide

Leak Alert Troubleshooting

One beep, one solid yellow light; check Visicheck feature for proper application.

Canister Full Alert Fixes

Two beeps, one solid yellow light; check canister, tubing, or call for service.

Low Battery Alert Indicators

Beeps and yellow lights indicate low battery; replace AA batteries.

System Error Alert Resolution

Repeating beep, solid yellow lights indicate system error; troubleshoot or call.

Device Lifecycle Expired Notification

Three lights and beep signal end of life; unit shuts off.

Blockage Alert Detection

Audible beeps and yellow light indicate blockage; check canister, tubing, clamp.

Alert Mute Button Functionality

Press and hold mute button to silence audible alerts for two minutes.

Therapy Ended Status

Solid yellow LED, beeps indicate therapy completion; unit will turn off.

Overview

The 3M Prevena™ 125 Therapy Unit is a medical device designed for incision therapy, utilizing negative pressure wound therapy (NPWT) to manage surgical incisions. It is a portable unit that provides continuous therapy and features various indicators and alerts to ensure proper operation and patient safety.

Function Description

The primary function of the Prevena 125 Therapy Unit is to apply negative pressure to surgical incisions, which is believed to help reduce edema, improve perfusion, and promote healing. The device creates a sealed environment over the incision using a specialized dressing, and then applies controlled negative pressure to the area.

Important Technical Specifications

  • Power Source: The therapy unit is powered by three AA batteries. Lithium batteries are recommended for optimal performance.
  • Therapy Duration (Lifecycle): The device has a lifecycle of 192 hours (eight days) of continuous therapy. After this period, the unit will shut off.
  • Battery Life Indicators: The unit provides visual indicators for battery charge levels:
    • Full charge: Approximately 9 hours of therapy remaining.
    • Medium charge: Approximately 2-7 hours of therapy remaining.
    • Low charge: Approximately 2 hours of therapy remaining, requiring immediate charging to avoid therapy disruption.
  • Audible Alarms: The device incorporates various audible alarms to signal different alert conditions, such as leaks, low battery, canister full, blockages, and system errors. These alarms can be muted temporarily.
  • Visual Indicators (LEDs): The unit uses solid yellow lights and flashing LEDs to communicate device status and alert conditions.
  • Visicheck™ Feature: This feature helps ensure proper application of the 3M Prevena Incision Dressing by indicating the leak rate condition:
    • One light: Best
    • Two lights: Good
    • Three lights: Marginal

Usage Features

  • Ease of Application: The device is designed for use with 3M Prevena Incision Dressings. The Visicheck™ Feature guides users in achieving an optimal seal.
  • Canister Management: The unit uses a canister to collect exudate. Users need to ensure the canister is securely locked into the unit, indicated by a distinct click.
  • Tubing Management: Proper connection and positioning of the tubing are crucial. Users should check for kinks, pinched tubing, and ensure connectors are fully engaged.
  • On/Off Control: The unit can be turned on and off by pressing and holding the ON/OFF button for specific durations (e.g., five seconds to turn off, two seconds to turn on).
  • Alert Mute Button: A dedicated button allows users to temporarily silence audible alerts for two minutes. The button illuminates when mute is active. Alerts will re-occur if the condition is not corrected within two minutes.
  • Battery Replacement: The battery panel on the back of the unit can be opened by pressing and pushing, allowing for easy replacement of the three AA batteries.
  • Therapy Completion Indicator: A solid yellow LED at the top of the Therapy Life Indicator signals the end of the therapy cycle. The unit will sound eight beeps, followed by a continuous beep for five seconds, and then shut off.

Maintenance Features

  • Leak Troubleshooting: If a "Marginal" leak rate is indicated by the Visicheck™ Feature, users are guided to:
    • Ensure the canister is locked.
    • Check tubing connectors for engagement.
    • Smooth out creases in the adhesive.
    • Use Prevena Incision Dressing Patch Strips to help seal leaks by pressing firmly around dressing edges.
  • Canister Full Troubleshooting: If the canister full alert sounds, users should:
    • Check for kinks or pinched tubing if the canister isn't full.
    • Turn the unit off if the canister is full.
    • Contact customer support (1-800-275-4524) for further assistance.
  • Low Battery Management: The unit provides clear visual and audible alerts for low battery. Users are advised to charge the unit immediately when a low charge is indicated to prevent therapy disruption.
  • System Error Resolution: In case of a system error, users are instructed to turn the unit off for five seconds, then turn it back on for two seconds. If the therapy is still not running, customer support (1-800-275-4524) should be contacted.
  • Device Lifecycle Management: The device has a finite lifecycle of 192 hours. Upon completion, the unit will shut off. If an attempt is made to turn it on after the lifecycle has expired, it will sound an alert for three seconds and then shut off. This indicates the need for device replacement.
  • Customer Support: A dedicated phone number (1-800-275-4524) is provided for NPWT product support, and another (1-800-228-3957) for general inquiries, along with the 3M medical website (3M.com/medical).