3
Ranger
Model 245
Blood/Fluid Warming System
For I.V. Use
3
Ranger
Model 247
Irrigation Fluid Warming System
Not for I.V. Use
1  Model 245 • Blood/Fluid Warming System
• For I.V. Use
1  Model 247 • Irrigation Fluid Warming System
• Not for I.V. Use
3  Modell 245 • Blut- und Infusionswärmesystem
• Für intravenöse Anwendungen
3  Modell 247 • Slflüssigkeit-Erwärmungssystem
• Nicht zum intravenösen Gebrauch
2  Modèle 245 • Système de réchauffement de
sang/soluté • Pour perfusion I.V
2  Modèle 247 • Système de réchauffement de liquide
d’irrigation • Ne pas utiliser pour une perfusion I.V
Service Manual 2
Manuel dentretien 23
Service-Handbuch 45

Other manuals for 3M Ranger 245

Questions and Answers

  • R
    Robert AyalaSep 10, 2025
    What causes a 3M Laboratory Equipment alarm to sound even after the unit is turned off?
    • C
      Charles AndrewsSep 10, 2025
      If the alarm on your 3M Laboratory Equipment continues to sound even after the unit has been turned off, it indicates that the independent backup safety system has been activated or the over-temperature screw is loose. Unplug the unit and make sure the over-temperature screw is completely tightened. If the screw is missing, turn the unit off and unplug it.
  • C
    Christine LloydAug 23, 2025
    What to do if my 3M Ranger 245 Medical Equipment alerts sound but unit has been turned off?
    • S
      Stacy TuckerAug 23, 2025
      If the 3M Medical Equipment continues to alert even after being turned off, it means that the independent backup safety system has been activated. Unplug the unit. Also, ensure the over-temperature screw is completely tightened. If it is missing, turn the unit off, unplug it.
  • L
    Luke TaylorAug 25, 2025
    What to do if 3M Ranger 245 alerts sound, alphanumeric display and alert indicator light go dark?
    • W
      wallmarkAug 25, 2025
      If the alert sounds, the alphanumeric display and alert indicator light go dark on your 3M Medical Equipment, this indicates a primary controller failure, and the unit will no longer operate. Turn the unit off and unplug it. Discontinue use of the unit and discard the warming set. Note that the alert will continue to sound if you do not unplug the unit.
  • G
    George JacksonAug 27, 2025
    Why nothing illuminates on the 3M Medical Equipment warming unit panel?
    • M
      Mary BrownAug 27, 2025
      If nothing illuminates on the warming unit panel of your 3M Medical Equipment, it could be due to several reasons: * The unit is not turned on. * The unit is not plugged in, or the power cord is not correctly plugged into an appropriate outlet. Ensure the power cord is plugged into the power entry module of the warming unit. * There may be a unit failure. Check the fuses.
  • C
    Chris BaileyAug 27, 2025
    What to do if my 3M Medical Equipment alerts soon after plugging it in?
    • A
      Adrian HerreraAug 28, 2025
      If your 3M Medical Equipment alerts soon after plugging it in, it might be due to a loose or missing over-temperature screw on the bottom of the unit. Make sure the over-temperature screw is completely tightened. If it is missing, turn the unit off and unplug it.
  • B
    Barbara LopezAug 29, 2025
    How to remove cassette from 3M Ranger 245 Medical Equipment?
    • D
      Diane AveryAug 29, 2025
      To remove the cassette from the 3M Medical Equipment, ensure that: * Fluid is drained from the cassette before sliding it out. * Fluids are no longer infusing. * The clamp is closed proximal to the warming cassette. Also, if the warming unit is below the patient level, raise it above the patient level to avoid excessive back pressure.
  • J
    James BowmanSep 18, 2025
    What does 'Er 5' or 'Open' mean on my 3M Laboratory Equipment?
    • D
      Dr. Jennifer Mathis DDSSep 18, 2025
      If the alphanumeric display reads 'Er 5' or 'Open' on your 3M Laboratory Equipment, it indicates electrical interference. Remove the unit.
  • G
    gina19Sep 12, 2025
    What does it mean when my 3M Laboratory Equipment displays a 'LO' temperature alarm?
    • G
      Gregory HarrisSep 12, 2025
      If your 3M Laboratory Equipment displays a 'LO' temperature alarm, it may be due to high flow using very cold fluid or a defective heater/relay. The alarm should stop when the temperature rises above 33°C. If the alarm continues, unplug the unit and discontinue use.
  • L
    Lisa ParkSep 8, 2025
    Why does my 3M Laboratory Equipment alarm right after plugging it in?
    • M
      michael92Sep 8, 2025
      If your 3M Laboratory Equipment alarms soon after being plugged in, check the over-temperature screw on the bottom of the unit. Ensure it is completely tightened. If the screw is missing, turn the unit off and unplug it.
  • J
    James FletcherSep 5, 2025
    How to fix a 3M Laboratory Equipment warming unit when nothing illuminates on the panel?
    • R
      Robert MannSep 5, 2025
      If nothing illuminates on the warming unit panel, first, ensure the unit is turned on and properly plugged into both the power entry module and a grounded outlet. Then check the fuses. If the issue persists, it may indicate a unit failure.

Summary

Technical Service and Order Placement

Technical Service Contacts

Contact information for technical support and service inquiries for Arizant products.

Order Placement Information

Information and contact details for ordering Arizant products and replacement parts.

In-warranty Repair and Exchange Process

Process for returning devices for service under warranty, including obtaining RA numbers.

Technical Support Call Requirements

Information required when contacting technical support, such as serial number.

Product Description and Overview

Ranger Warming Unit Features

Details the features and components of the Ranger warming unit.

Maintenance and Storage Procedures

Cleaning the Ranger Warming Unit

Instructions on how to clean the Ranger warming unit, including precautions.

Cleaning the Exterior of the Warming Unit

Specific steps for wiping down the outside of the warming unit.

Cleaning the Heating Plates

Procedures for cleaning the internal heating plates of the warming unit.

Storage Guidelines

Guidelines for storing the warming unit components when not in use.

Servicing Information

Information regarding user-serviceable parts and authorized service requirements.

Model 245 Preventative Maintenance

Tools and Equipment for Model 245 PM

Lists necessary tools and equipment for performing preventative maintenance on the Model 245.

Model 247 Preventative Maintenance

Tools and Equipment for Model 247 PM

Lists necessary tools and equipment for performing preventative maintenance on the Model 247.

Replacing the Power Fuse

Tools and Equipment for Fuse Replacement

Lists the small slotted screwdriver needed for fuse replacement.

Method for Fuse Replacement

Step-by-step instructions for safely replacing the power fuse in the unit.