Ranger Blood and Fluid Warming Systems
English
8
Arizant Healthcare Inc.
Maintenance and Storage
Cleaning the RangeR WaRming unit
Clean the Ranger warming unit on an as-needed basis�
caution
• Do not immerse the warming unit in cleaning or sterilizing solutions� e unit is not liquid-proof�
• Do not clean the warming unit with solvents or damage to the case, label, and internal components
may result�
to Clean the exteRioR of the WaRming unit:
1� Disconnect the Ranger warming unit from the power source�
2� Wipe the outside of the unit with warm, soapy water, nonabrasive cleaning solutions, dilute
bleach, or cold sterilants� Do not use abrasive materials�
3� Wipe with a dry, so cloth�
to Clean the heating Plates:
e Ranger cleaning tool (part #90030) is intended to clean both heating plates of the warming unit�
It is not necessary to disassemble the warming unit to use the tool�
caution
• Do not insert metallic instruments in the warming unit, as damage to the heating plates may
occur�
• Do not use abrasive material or solutions to clean the heater plates�
• Do not allow spills to dry inside the unit, as this may make it more dicult to clean the unit�
• e cleaning tool provides only supercial cleaning—it does not disinfect or sterilize the interior of
the unit�