Ranger Model 247 Maintenance Log
The operating temperature and alarms should be tested at least every 6 months or according
to institutional protocol.
maintenanCe aCtion PeRfoRmed
model ___________ sn _______
Date ________ Operating temperature: (Specication = 41°1) ______°C
Over-temperature alarm points:
Audible (48°C [±2] on Ranger unit display) ______°C
Secondary (50°C [±2]) ______°C
Date ________ Operating temperature: (Specication = 41°1) ______°C
Over-temperature alarm points:
Audible (48°C [±2] on Ranger unit display) ______°C
Secondary (50°C [±2]) ______°C
Date ________ Operating temperature: (Specication = 41°1) ______°C
Over-temperature alarm points:
Audible (48°C [±2] on Ranger unit display) ______°C
Secondary (50°C [±2]) ______°C
Date ________ Operating temperature: (Specication = 41°1) ______°C
Over-temperature alarm points:
Audible (48°C [±2] on Ranger unit display) ______°C
Secondary (50°C [±2]) ______°C
Date ________ Operating temperature: (Specication = 41°1) ______°C
Over-temperature alarm points:
Audible (48°C [±2] on Ranger unit display) ______°C
Secondary (50°C [±2]) ______°C
Date ________ Operating temperature: (Specication = 41°1) ______°C
Over-temperature alarm points:
Audible (48°C [±2] on Ranger unit display) ______°C
Secondary (50°C [±2]) ______°C
Model 247