Ranger Blood and Fluid Warming Systems
English
20
Arizant Healthcare Inc.
Specifications
PhysiCal ChaRaCteRistiCs
Warming unit
4�5 in� (11 cm) high x 7�5 in� (19 cm) wide x 10 in� (25 cm) long;
wt�: 7 lb� 4 oz� (3�3 kg)
claSSiFication
Classied under IEC 601-1 Guidelines (and other national versions of the Guidelines) as Class I, Type
B, Ordinary Equipment� Not suitable for use in the presence of ammable anesthetic mixtures with
air or with oxygen or nitrous oxide� Classied by Underwriters Laboratories Inc� with respect to
electric shock, re, and mechanical hazards only, in accordance with UL 60601-1 and in accordance
with Canadian/CSA C22�2 No� 601�1�
eleCtRiCal ChaRaCteRistiCs temPeRatuRe ChaRaCteRistiCs
input voltage Set point temperature
100-120 or 220-240 VAC 41°C (±1°C)
operating Frequency over-temperature alarm
100-120 VAC, 50/60 Hz Model 245: 43°C (±1°C)
220-240 VAC, 50/60 Hz Model 247: 48°C (±2°C)
maximum heating poWer under-temperature alarm
900 W 33°C
FuSe over-temperature cutoFF
2 x 10 Amp for 100 – 120 VAC Model 245: 46°C (+3, -2°C)
2 x 6�3 Amp for 220 – 240 VAC Model 247: 50°C (±2°C)
leakage CuRRent
Meets leakage current requirements in accordance with UL 60601-1 and IEC 60601-1�
enviRonmental Conditions
operating temperature range
15° to 40°C (59° to 104°F)
Storage temperature range
-20° to 45°C (-4° to 113°F)
operating humidity
10 to 85% RH, noncondensing
atmoSpheric preSSure range
50 kPa to 106 kPa