Ranger Blood and Fluid Warming Systems
Ranger Model 245 Maintenance Log
The operating temperature and alarms should be tested at least every 6 months or according
to institutional protocol.
maintenanCe aCtion PeRfoRmed
model ___________ sn ________
Date _________ Operating temperature: (Specication = 41°C±1) ______°C
Over-temperature alarm points:
Audible (43°C [±1] on Ranger unit display) ______°C
Secondary (46°C [+3, -2]) ______°C
Date _________ Operating temperature: (Specication = 41°C±1) ______°C
Over-temperature alarm points:
Audible (43°C [±1] on Ranger unit display) ______°C
Secondary (46°C [+3, -2]) ______°C
Date _________ Operating temperature: (Specication = 41°C±1) ______°C
Over-temperature alarm points:
Audible (43°C [±1] on Ranger unit display) ______°C
Secondary (46°C [+3, -2]) ______°C
Date _________ Operating temperature: (Specication = 41°C±1) ______°C
Over-temperature alarm points:
Audible (43°C [±1] on Ranger unit display) ______°C
Secondary (46°C [+3, -2]) ______°C
Date _________ Operating temperature: (Specication = 41°C±1) ______°C
Over-temperature alarm points:
Audible (43°C [±1] on Ranger unit display) ______°C
Secondary (46°C [+3, -2]) ______°C
Date _________ Operating temperature: (Specication = 41°C±1) ______°C
Over-temperature alarm points:
Audible (43°C [±1] on Ranger unit display) ______°C
Secondary (46°C [+3, -2]) ______°C
Model 245