EasyManua.ls Logo

3M Ranger 245 - Page 12

3M Ranger 245
22 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Ranger Blood and Fluid Warming Systems
English
12
Arizant Healthcare Inc.
testing the oveR-temPeRatuRe alaRm Points (43°C [±1] and 46°C [+3, -2])
e over-temperature alarm test causes the Ranger unit to go into an over-temperature test mode by
manually overriding the primary control system and engaging the heaters� e Ranger system is very
responsive to heater input; therefore the test procedures are technique sensitive� Read instructions
thoroughly before beginning these tests�
e rst alarm point at 43°C (±1) alerts user to the rise in temperature (see step 7)� At the
46°C (+3, -2) secondary alarm set point, the unit cuts power to the heaters (see step 8)� Please note that
due to the heaters’ rapid response, you may see temperature readings dri within the range of
44-49°C
7 Loosen the over-temperature screw on the
bottom of the Ranger unit (see Fig� 3)� Listen
for a slight click as the microswitch opens�
Loosen ½ turn more. is unit is now
in over-temperature test mode, and the
temperature output of the unit should begin
to rise� When the Ranger display reads 43°C,
verify that an audible alarm sounds, the alarm
light illuminates, and the display alternately
ashes “HI” and the plate temperature�
8� Listen for the power relay to click at
46°C (+3, -2) (you can feel the click when holding the lower le corner of the unit as you face
the front display)� e click signies the secondary alarm trip point and signals that power to
the heaters has been shut o� e over-temperature alarm sounds, the alarm light illuminates,
and the alphanumeric display alternately reads “HI” and the plate temperature� Watch the
thermocouple temperature monitor to determine the units peak temperature output (less than
1 minute)� e temperature reading on the monitor should peak at 46°C (+3, -2)
note: If the temperature displayed on the thermocouple monitor exceeds 49°C during the test, tighten
the over-temperature test screw on the bottom of the Ranger unit, unplug it, and run cold uids
through the warming cassette� is returns the unit to normal operating temperature� Call
Arizant Healthcare Inc� Technical Support
9 e secondary alarm is designed to latch� Verify this feature by momentarily turning o the
power switch, then turning it on again� e Ranger unit should continue alarming throughout
this sequence�
10� Securely tighten the over-temperature screw, turn o the Ranger unit, and unplug the unit�
Once power is cut to the Ranger unit, the alarm will cease� e unit is ready to return to service
aer it cools to normal operating temperature (approximately 20-30 minutes)�
Fig. 3: Ranger over-temperature screw
Over-temperature
screw
Model 245

Other manuals for 3M Ranger 245

Related product manuals