English
17
247021E English
Ranger Blood and Fluid Warming Systems
Replacing the Power Fuse
e power fuses are located in the power entry module�
tools and equiPment
Small slotted screwdriver
method
1� Disconnect the warming unit from the power source�
2� Remove the power cord�
3� Locate the fuse carrier in the power entry module�
4� Use the small screwdriver to remove the fuse carrier from the power entry module�
5� Remove the blown fuse from the fuse carrier and replace with a new fuse (as marked)�
6� Place the fuse carrier back into the power entry module�
7� Reattach the power cord�
8� Reconnect the warming unit to the power source�
9� Turn the unit on to verify normal operation�
10� Record the maintenance action taken�