Change the measurement frequency and offset by OK key
The results of different frequency measurements are different, so it
needs set to the specified frequency.
Different frequency measurement results will have differences, so it is
basically set to the specified frequency. "offset" is used when correcting
errors or converting external attenuators or amplifiers. For example, the
signal enters the instrument through a 30dB attenuator. The offset is set
to 30dB. If a 30dB amplifier enters the instrument, the offset is set to
-30dB.This parameter can also be set to compensate if errors are found in
the measurement of the instrument itself.
"Average power (dBm)" is the dBm readings of the average after
measuring 200 times. "Average power" is the reading when the reading is
converted to MW. "Average power" is the reading when the reading is
converted to W.
The power hold: it is keeping display measured the average power
value, so it is easy to capture the pulse power signal.
Maximum power (dBm) "is to take the largest value of all measured
values ( dBm) reading."Maximum power (mW)" is the readings converted
to mW. Click the OK key to reset the maximum value.