Physical Characteristics
Outline Dimensions See Figures 2-3 through 2-6
Vibration Limits 1.5 g for 10 - 150 Hz
Signal Cable
(supplied with the
signal converter)
Non-Permanently Installed Cable:
100 ft. (33 m) Max.
Permanently Installed Cable:
Standard Length - 50 feet (16 m)
Optional Length - Up to 100 feet (33 m) in 10 foot
increments, as specified
Materials of
Construction
Meter Liner see Section 1.2 Model Number Breakdown
Electrodes see Section 1.2 Model Number Breakdown
Meter Body 304 SST, epoxy finish
Flanges carbon steel or 304 SST, as specified
Meter Housing cast aluminum, epoxy finish
Electronics Housing cast aluminum, epoxy finish, 316 SST attachment screws,
gasketed cover
Meter Enclosure Ratings
NOTE
Enclosures are suitable for indoor or outdoor installation.
Enclosure ratings apply to the meter.
Watertight Housing
(standard)
NEMA 4X, IEC 529 IP65
Accidental
Submergence
NEMA 4X, IEC 529 IP67, 30 feet H
2
O/48 h (9 m H
2
O/48 h)
Continuous
Submergence
NEMA 4X, IEC 529 IP68, 30 feet H
2
O (9 m H
2
O)
Conduit Connections two 1/2 inch NPT internally threaded entrances
Certifications refer to Section 1.2 Model Number Breakdown
10DS3111E INSTRUCTION MANUAL
1-9