178  AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B 
 
 
 
Changing the unit 
1  Press COMP. CONFIG. softkey. The "Configure Component" menu is 
displayed with the component list. 
2  In the component list select the component with unit which is to be 
changed. 
 
 
 
3  Press UNIT SELECTION softkey. Enter password (level 3), as applicable. 
A list of the units available is displayed. 
4  Select the desired unit and confirm with ENTER. 
5  A security prompt appears with the information that when switching 
the unit, the range limits of the component are automatically adjusted. 
6  Confirm the switching of the unit with ENTER or cancel with Back. 
7  Quit the menu with Back or Meas. 
8  Check the range limits (see page 156) and the number of decimal places 
(see page 158) and adjust if necessary.