188  AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B 
 
 
 
Serial number 
The 14-digit serial number of the analyzer modules is found in the analyzer 
data sheet and on a label affixed to the module, generally on the CPU board. 
The serial number contains the following information (for example): 
01400000012301 
The first 3 digits indicate the module type: 
014 
Analyzer module 
004 
Analyzer module 
006 
Cooler I/O board (installed in the SCC-F Sample gas feed unit as an 
option) 
008 
LS25 Laser analyzer module 
The remaining 11 digits are the actual serial number of the module. 
Function block application 
When setting up the digital I/O module, a function block application must be 
configured. It is also possible to assign another function block application to 
this system module during operation. The standard function block applica-
tions with the respective connection layouts of the module are listed in the 
"Electrical connections: Digital I/O module" (see page 112) section. 
Unknown system module 
There are several reasons for a system module to have an Unknown status 
in the Setup system modules menu item: 
Cause: The system module was not found after the power supply was turned 
on (status message no. 201).   
Corrective action Reestablish the system bus connection to the system 
module and press the RESTART softkey. 
Cause: The system bus connection to the system module is broken (status 
message no. 209).   
Corrective action: Reestablish the system bus connection to the system 
module and press the RESTART softkey. 
Cause: The system module serial number was entered incorrectly.   
Corrective action: Press the CHANGE softkey and correct the serial number. 
NOTE 
Automatic calibration of an analyzer module is not possible during setting 
up system modules.