AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B  257
 
 
 
Calibration data for an analyzer module with 1 sample component 
Example: Test gas = air 
Calibration data for manual calibration 
Calibration method  Common single-point calibration 
Calibration measurement range  0 to 25 vol.% of O
2
 
Test gas concentration  20.96 vol.% of O
2
 
   
Calibration data for automatic calibration 
Calibration method  Common calibration (test gas) 
Single zero-point calibration  Always 
Single span calibration  Never 
Common zero and span calibration  Never 
Test gas concentration zero gas  20.96 vol.% of O
2
 
Test gas concentration span gas  n.a. 
Calibration data for an analyzer module with > 1 sample component 
Example: Test gas = air 
Calibration data for manual calibration 
Calibration method  Substitute gas calibration 
Zero component  O
2
 in N
2
 
Zero range  0 to 25 vol.% of O
2
 
Span component  n.a. 
Span range  n.a. 
Test gas concentration zero gas  20.96 vol.% of O
2
 
Test gas concentration span gas  n.a. 
 
Calibration data for automatic calibration 
Calibration method  Substitute gas calibration
 1)
 
Single zero-point calibration  Always 
Single span calibration  Never 
Common zero and span calibration  Never 
Test gas concentration zero gas  20.96 vol.% of O
2
 
Test gas concentration span gas  n.a. 
1)  Settings of sample components and measurement ranges for zero and 
span calibration are taken over from the calibration data for manual 
calibration.