AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B  43
 
 
 
Module power consumption 
 
Module Power consumption 
System controller  approx. 15 W 
I/O modules  each approx. 10 W 
Caldos25  max. 25 W 
Caldos27  max. 17 W 
Fidas24  max. 40 W 
Fidas24 NMHC  max. 40 W 
Limas11 IR  max. 100 W 
Limas21 UV  max. 100 W 
Limas21 HW  max. 100 W 
Magnos206  max. 50 W 
Magnos28  max. 50 W 
Magnos27  max. 35 W 
Uras26  max. 95 W 
ZO23  approx. 12/35 W in continuous/starting operation 
Pneumatic module    approx. 20 W 
Fidas24, Fidas24 NMHC: Detector and sample gas inlet heaters 
 
Input voltage  115 VAC or 230 VAC, ± 15 % (max. 250 VAC) 
Line frequency range  47 to 63 Hz 
Power consumption  125 VA for detector Fidas24,   
approx. 200 VA for detector Fidas24 NMHC,   
125 VA for sample gas inlet (option) 
Connection  4-pin connector, connecting cable supplied 
Safety 
 
Test  to EN 61010-1:2010 
Class of protection  Central unit with electronics module (power sup-
ply): I; Analyzer module without electronics mod-
ule (power supply): III 
Overvoltage category /   
degree of pollution 
Power supply: II/2 
Safe isolation  The electronics module power supply is galvani-
cally isolated from other circuits by means of 
reinforced or double insulation. Functional ex-
tra-low voltage (PELV) on the low-voltage side. 
Electromagnetic compatibility 
 
Interference immunity  Tested to EN 61326-1:2013. Inspection severity: 
Industrial area, fulfills at least the rating to Table 2 
of EN 61326. 
Emitted interference  Tested to EN 61326-1:2013. Limit value class B for 
interference field strength and interference volt-
age is met.