CoriolisMaster FCB400, FCH400 CORIOLIS MASS FLOWMETER | OI/FCB400/FCH400-EN REV. E 121
Software history
In accordance with NAMUR recommendation NE53, ABB offers a transparent and traceable software history.
Device software package FCx4xx (Device Firmware Package)
Version Issue date Type of change Description Ordering number
00.01.00 6/12/2015 New release 3KXF002043U0100
00.03.00 01.2016 Change SIL2 Functionality and minor troubleshooting 3KXF002043U0100
00.04.00 11.2016 Change Variety of minor changes, details in document MI/FCX400/FW/101
01.2017
3KXF002043U01000
00:05:00 01.2018 Change Variety of minor changes, details in document MI/FCX400/FW/102
01.2018
3KXF002043U0100
Change from one to tw o columns
Zero point balance under operating
conditions
Devices in the CoriolisMaster series do not necessarily require
zero point adjustment. Performing a zero point adjustment is
only recommended in the following cases:
For measurements in the lower flow range
(below 10 % of Q
max
DN).
If particularly high accuracies are required (0.1 % or
better).
If the operating conditions (pressure and temperature)
deviate greatly from the reference conditions (see data
sheet).
For zero point adjustment under operating conditions, make
sure the following conditions are present:
The meter tube is completely filled with the measuring
medium.
For liquid measuring media, no gas bubbles or air
pockets may be present in the meter tube.
For gaseous measuring media, no liquid components or
condensates may be present in the meter tube.
The pressure and the temperature in the meter tube
correspond to standard operating conditions and are
stable.
In case of an increased zero point (> 0.1 %), check the installation
for ‘best praxis’ and make sure that no gas content is contained
in liquids, or that there are no liquids or particles in gases.
See also Turn-off devices for the zero point adjustment on
page 33.
Zero point balancing can be started either manually via the LCD
indicator or using an appropriately configured digital input.
See Menu: Device Setup on page 99 and Menu: Input / Output
on page 107.