ProcessMaster FEP610, HygienicMaster FEH610 | OI/FEP610/FEH610-EN Rev. B  41 
 
 
7.3.3  Error messages on the LCD display 
In the event of an error, a message consisting of a symbol and 
text (e.g. Electronics) appears at the bottom of the process 
screen.  
The text displayed provides information about the area in 
which the error has occurred. 
 
Process display 
 
 
 
 
 
 
 Electronics           
 
The error messages are divided into four groups in 
accordance with the NAMUR classification scheme. The group 
assignment can only be changed using a DTM or EDD: 
 
Symbol  Description 
 
Error / failure 
 
 
Function check 
 
 
Outside of the specification 
 
 
Maintenance required 
 
 
The error messages are also divided into the following areas: 
 
Range 
Description 
Operation  Error / alarm due to the current operating 
conditions. 
Sensor  Error / alarm of the flowmeter sensor. 
Electronics  Error / alarm of the electronics. 
Configuration  Error / alarm due to device configuration. 
 
NOTE  
For a detailed description of errors and troubleshooting 
instructions, please see chapter " 
 
Diagnosis / error messages" on page 64. 
 
 
 
7.3.4  Switching to the configuration level 
(parameterization) 
The device parameters can be displayed and changed on the 
configuration level. 
 
Process display 
 
 
 
 
 
 
 
 
1.  Use   to switch to the configuration level. 
 
ccess Level 
 
 
Read Only 
Standard 
Advanced 
Back  Select 
 
2.  Select the desired level of access using   /  . 
3.  Confirm the selection with  . 
 
NOTE
  
There are three levels of access. A password can be defined 
for level "Standard".  
There is no factory default password. 
 
Access Level  Description 
Read Only  All parameters are locked. Parameters are read 
only and cannot be modified. 
Standard  All the parameters can be altered. 
Service  Only Customer Service has access to the Service 
menu.