Type | Electromagnetic Flowmeter |
---|---|
Measurement Principle | Faraday's Law of Electromagnetic Induction |
Process connection | Flanged, Wafer, Threaded, Sanitary |
Flow Tube Material | Stainless steel, Hastelloy C |
Liner Options | PTFE, PFA, Polyurethane, Hard Rubber |
Electrode Options | Stainless steel, Hastelloy, Tantalum, Platinum |
Accuracy | ±0.2% of reading |
Repeatability | ±0.1% of reading |
Fluid temperature | -40°C to +180°C (depending on liner) |
Ambient temperature | -40°C to 60°C |
Protection class | IP67, IP68 |
Output signal | 4-20 mA, HART |
Certifications | ATEX, IECEx, FM, CSA |
Sizes | DN 15 to DN 2000 |
Details the purpose of the MagMaster to WaterMaster electronics upgrade and lists essential related publications.
Explains DIP switch functions (SW1, SW2, SW3) on the transmitter backplane for MagMaster retrofit.
Provides instructions for correctly wiring the MagMaster sensor cable to the WaterMaster transmitter.
Specifies the minimum 15 Ω sensor coil resistance needed for WaterMaster transmitter operation.
Covers the initial system startup message and how to access the WaterMaster programming menus.
Details key 'Sensor Setup' parameters (Type, Size, Cal Mode, Span, Zero) for retrofitted MagMaster sensors.
Guides on entering calibration dates, locations, correction factors, and setting the final 'Calibrated' status.