User Guide IM/MM2WM–EN Rev. A
MagMaster
Electromagnetic flowmeter
MagMaster to WaterMaster
electronics upgrade
Introduction
These instructions describe how to upgrade an existing
MagMaster MF/F flowmeter (DN80 to DN1600)
installation with remote transmitter for use with the new
remote WaterMaster transmitter.
This guide should be read in conjunction with the
following publications:
– WaterMaster FET100 User Guide, OI/FET100–EN
– WaterMaster Programming Guide, IM/WMP
SP033 Adam Street Wynnum SPS - WaterMaster - Electromagnetic Flowmeter -
OIML R49 Class 1 450mm - User Guide
Q-Pulse Id: VM486
Active: 08/04/2016
Page 1 of 12
Question and Answer IconNeed help?

Do you have a question about the ABB MagMaster MF/F and is the answer not in the manual?

ABB MagMaster MF/F Specifications

General IconGeneral
TypeElectromagnetic Flowmeter
Measurement PrincipleFaraday's Law of Electromagnetic Induction
Process connectionFlanged, Wafer, Threaded, Sanitary
Flow Tube MaterialStainless steel, Hastelloy C
Liner OptionsPTFE, PFA, Polyurethane, Hard Rubber
Electrode OptionsStainless steel, Hastelloy, Tantalum, Platinum
Accuracy±0.2% of reading
Repeatability±0.1% of reading
Fluid temperature-40°C to +180°C (depending on liner)
Ambient temperature-40°C to 60°C
Protection classIP67, IP68
Output signal4-20 mA, HART
CertificationsATEX, IECEx, FM, CSA
SizesDN 15 to DN 2000

Summary

MagMaster to WaterMaster Electronics Upgrade Introduction

Upgrade Purpose and Related Documents

Details the purpose of the MagMaster to WaterMaster electronics upgrade and lists essential related publications.

Electrical Installation and DIP Switch Configuration

Configuring MagMaster Transmitter DIP Switches

Explains DIP switch functions (SW1, SW2, SW3) on the transmitter backplane for MagMaster retrofit.

Sensor Cable Wiring and Coil Resistance

MagMaster Sensor Cable Connection Guide

Provides instructions for correctly wiring the MagMaster sensor cable to the WaterMaster transmitter.

WaterMaster Sensor Coil Resistance Requirement

Specifies the minimum 15 Ω sensor coil resistance needed for WaterMaster transmitter operation.

Programming and Sensor Setup Procedures

Initial Startup and Menu Navigation

Covers the initial system startup message and how to access the WaterMaster programming menus.

Configuring Sensor Setup Parameters

Details key 'Sensor Setup' parameters (Type, Size, Cal Mode, Span, Zero) for retrofitted MagMaster sensors.

Calibration Entry and Final Status Setting

Guides on entering calibration dates, locations, correction factors, and setting the final 'Calibrated' status.