EasyManua.ls Logo

ABB S4Cplus

ABB S4Cplus
86 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3UHSDULQJWKHHTXLSPHQW
2.2. Start up of Levelmeter 2000
313HAC 16578-1 Revision: E
© Copyright 2003-2007 ABB. All rights reserved.
3RZHUVRXUFHVRI/HYHOPHWHU
Two alternatives are available:
Battery mode: turn on the Levelmeter by pressing ON/MODE until display flashes.
This will eliminate battery saving mode. Do not forget to turn off after use.
External power: connect the power cords (red/black) to 12-48 VDC, found in the
cabinet or external transformer.
$GGUHVV
,GHQWLI\VHQVRUV
5HVXOW
The Levelmeter 2000 is now initialized and ready for service.
4. 0 connect the Levelmeter 2000 to the
COM2 port in the control cabinet through the
connector marked 287FRQQHFWLRQ6,2.
0 connect the Levelmeter 2000 to the
COM1 port in the control cabinet through the
connector marked 287FRQQHFWLRQ6,2.
Shown in the figure /D\RXWDQG
FRQQHFWLRQRI/HYHOPHWHURQSDJH
.
5. Calibrate the robot!
$FWLRQ 1RWH
$FWLRQ 1RWH
1. Make sure the sensors have different addresses. Any
addresses will do, as long as they differ from each
other.
Detailed in the documentation
supplied by sensor manufacturer,
found in the calibration tool set.
$FWLRQ 1RWH
1. Connect the sensor to the VHQVRUFRQQHFWLRQ points. Marked $ and %.
Shown in the figure /D\RXWDQG
FRQQHFWLRQRI/HYHOPHWHU
RQSDJH.
2. Press 2102'(.
3. Press 2102'( until the dot flashes under 6(1625.
4. Press ENTER.
5. Press =(526(/(&7 arrows until a flashing "A B" is
shown.
6. Press (17(5
Wait until the "A B" flashes again.
7. Press (17(5
&RQWLQXHG

Other manuals for ABB S4Cplus

Related product manuals