EasyManua.ls Logo

ABB WaterMaster - Specification - Sensor; Pressure limitations; Temperature limitations; Environmental protection

ABB WaterMaster
48 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
WaterMaster
Electromagnetic flowmeter 7 Specification – Sensor
IM/WM–EN Rev. M 39
7 Specification – Sensor
Functional specification
Pressure limitations
As per flange rating – non approved
PN16 for OIML R49 Approved
Temperature limitations
Environmental protection
Rating:
IP68 (NEMA 6) to 10m (33 ft) depth (not DN10 to DN32)
IP67 (NEMA 4X) – DN10 to DN32
Buriable (sensor only)
FEW (DN10 to 32) No
FEV and FEF Yes
Conductivity
>5µS cm
–1
Transmitter mounting
Integral or remote
Electrical connections
20 mm glands
1
/2 in NPT
20 mm armored glands
Sensor cable
ABB WaterMaster cable available in two forms –
standard and armored
Maximum length 200 m (660 ft)
Ambient temperature
Remote transmitter
Integral transmitter
–20 to 70 °C (–4 to 158 °F)
–20 to 60 °C (–4 to 140 °F)
Process temperature –6 to 70 °C (21 to 158 °F) – non approved
0.1 to 50 °C (32.2 to 122 °F) – OIML R49 T50
Approved

Other manuals for ABB WaterMaster