Appendix A: Product Specicaons
Storage and Excursion Temperature Specicaons
Store the components in this kit according to the following recommended condions.
Table 1. Recommended storage and excursion condions for components
Long Term Storage Temperature Store at 23°C or 73°F
Temperature Excursions Excursions below 15°C (59°F) or above 30°C (86°F)
should be <24 hours in duraon
Excursion Temperature Limits -20°C (-4°F) to 60°C to (140°F)
Product Materials
The following materials are intended to come into contact with ssue.
Table 2. Product materials for generator kit
Component Material
Generator Titanium, silicone rubber, epoxy resin
Pocket sizer Polybutylene terephthalate
Port plug Polyether ether ketone (PEEK) with barium sulfate
NOTE: These components are not made with natural rubber latex.
Generator Specicaons
The Eterna™ implantable pulse generator has the following physical specicaons.
Table 3. Generator specicaons
Specicaons Model 32400
MR Condional Yes
Upgradeable features Yes
BurstDR™
smulaon capable
Yes
Height 3.87 cm (1.52 in.)
Length 4.81 cm (1.89 in.)
Thickness 0.90 cm (0.36 in.)
Weight 26.4 g (0.93 oz)
Volume
13.79 cm
3
(0.84 in.
3
)
Power source Rechargeable lithium ion cell baery
Connector specicaons Two 8channel connectors
Inline, 2.54 mm (0.1 in.) spacing
Connector strength: 5N
Program storage capacity 16 programs with 8 sm sets each
11