Unpacking, Installaon & Transport
Unpack the Pima Analyser and place safely on a clean, at, level and stable
surface. Do not put the Pima Analyser on surfaces that may vibrate, e.g. the
top of a freezer or refrigerator or next to a laboratory centrifuge.
Inspect for any obvious sign of damage. Report any damage immediately to
your local distributor. Allow temperature equilibraon for at least 30 minutes
to ensure that the Pima Analyser has reached surrounding temperature.
When using the Pima Bag for transportaon, take out the Pima Analyser and
place safely on a clean, at, level and stable surface for tesng.
If the Pima Analyser was transported or moved to areas with dierent climac
condions, allow temperature equilibraon for at least 30 minutes.
Note: Keep the original Pima Analyser box in case the instrument needs to
be shipped for repair or the like.
Pima Analyser Calibraon
The Pima Analyser is factory calibrated and does not require any further
calibraon and vericaon operaons. However, a daily performance check
using Pima Bead Standard is recommended. If the Pima Analyser was
transported or moved, a test with the Pima Bead Standard may be run to ensure
proper funconality. Please refer to the Pima Bead Standard User Guide
(UG-PIMA-03) for more informaon on this product.
Pima Analyser Recharging Intervals during Storage
During long term storage below 30 °C, the Pima Analyser should be connected
to mains A/C power and the baery recharged overnight at least once per
year. During long term storage at 30 - 40 °C the baery needs to be recharged
at least once every 6 months. If long term storage temperatures exceed 40 °C,
the recharge intervals may need to be considerably shorter.
Note: It is not recommended to store the Pima Analyser with an empty
baery. Therefore, always recharge the baery before long term
storage.
1 INTRODUCTION

Questions and Answers: