EasyManuals Logo

aboni HydroTracer HT3 User Manual

aboni HydroTracer HT3
53 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #18 background imageLoading...
Page #18 background image
Performing a Test
16
2.5
Operation menu with manual input: Loading the HydroTracer
This chapter describes the steps of the manual operation as shown on the screen.
The images in the instructions window clarify the activity to be performed. Under the current
picture, the respective step is briefly described. The indicator to the right above the image
indicates the amount of time, in seconds, that has passed to complete the entire fill operation.
Press X to abort the test.
Note: The opening time should be at least 15 seconds but not more than 180 seconds.
Note: If you use the test pads instead of the powdered reagent, the procedure and the
number of instructions and their sequence are not exactly identical to the following
instructions. In principle, however, the performance of the filling process is comparable.
Open the test chamber by pulling the one-
hand lever up to the stop. The system fan
is switched off.
Remove the sample tray. Remove old
sample material. Place the sample tray on
the scale and tare it.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the aboni HydroTracer HT3 and is the answer not in the manual?

aboni HydroTracer HT3 Specifications

General IconGeneral
Brandaboni
ModelHydroTracer HT3
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals