68
The code chip is incorrect.
Turn off the meter and insert
a new code chip. If this does
not x the problem,
call Accu-Chek Customer
Care at 1-800-363-7949.
Your blood glucose may be
extremely high or a meter or a
test strip error has occurred.
If this conrms how you feel,
contact your healthcare
professional immediately. If
it does not conrm how you
feel, repeat the test and refer to
Chapter 2, “Unusual Blood Glucose Results.” If this
does not conrm the way you feel, run a control test
with your control solution and a new test strip. If the
control result is within the acceptable range, review
the proper testing procedure and repeat your blood
glucose test with a new test strip. If the E3 code
still appears for your blood glucose test, your blood
glucose result may be extremely high and above the
system’s reading range. Contact your healthcare
professional immediately. If the control result is
not within the acceptable range, see Chapter 4,
“Understanding Out-of-Range Control Test Results.”
43604_NanoAv_12hr_Ins_EN.indd 68 11/24/08 11:08:51 AM

Other manuals for Accu-Chek Aviva Nano

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Aviva Nano and is the answer not in the manual?

Accu-Chek Aviva Nano Specifications

General IconGeneral
Blood Sample Size0.6 µL
Test Time5 seconds
Battery LifeApproximately 1000 tests
Dimensions69 x 43 x 20 mm
DisplayLCD
Measuring Range10-600 mg/dL
Battery1 x 3V CR2032 lithium battery
Storage Conditions-25°C to 70°C
Memory500 test results
WeightApprox. 40 g (with battery)

Related product manuals