39
Chapter 1: Your New Meter
ACCU-CHEK Customer Service Center 1-800-858-8072
A test error has occurred.
Repeat the test. If occurs
again, please contact the ACCU-
CHEK Customer Care service center
at 1-800-858-8072.
The test drum in the meter is
empty. Insert a new drum. If the
message is still displayed, call
the ACCU-CHEK Customer Care
service center at
1-800-858-8072.
Note: A dot also appears in
the view window on the back
of the meter.
There was not enough blood on
the test strip to start the test.
Repeat the test with a larger
drop of blood.
Chapter 5: Maintenance and Troubleshooting ACCU-CHEK Customer Care Service Center 1-800-858-8072
The door was opened and closed
with a partly used test drum inside,
or a partly used test drum was
inserted.
You should run a control test (see
Why Run Control Tests?
) to verify
that the test drum is useable. The
flashing test drum symbol will still
be displayed until you insert a new
test drum.
mer Service Center 1-800-858-8072
are Service Center 1-800-858-8072
display check. If any parts
er segments are missing or
call the ACCU-CHEK
r Care service center at
58-8072.
perature is above or below
s of the meter (below 41
113°F). Move to an area
50 and 104°F, wait 30
or more and repeat the
r the meter has reached
temperature. Do not
y heat or cool the meter.
ding may not be accurate
the test was performed
he best operating
ture range of the system
°F). Results are displayed
ratures within 41 and 113°F.