PHSA Laboratories CW Site - Point of Care
Title: CWPC_GLU_0140 Accu-Chek Inform II Training Guide
Medical Approval: Dr Benjamin Jung Medical Approval Date: 27 Dec 2016
Version: 1.2 Implementation Date: 7/2/2019 11:55:59 PM
Folder Name: CW\Point of Care\Whole Blood Glucose
This is a controlled document for CW use only. Any printed copies are uncontrolled unless specified. Please refer to Lab QMS
Page 5 of 12
Hospital; Women’s Health Centre; Sunny Hill Health Centre; Canuck Place,
BC Cancer Agency – Vancouver Site).
Obtaining the Operator ID
Each Operator has a unique Operator ID # - DO NOT SHARE WITH OTHER
STAFF.
Bar Codes given to the Clinical Nursing Educator/Leader from Point of Care
Technologist.
(Quarterly Operator Status reports issued by POCT to the Clinical Nursing
Educator/ Leader)
Cleaning and Disinfecting the Meter
CAVIWIPES @ PHSA – this is the manufacturer approved cleaning wipe
DO NOT USE OTHER BRAND – eg. ACCEL TB wipes are corrosive to the meter
touch screen
Glucose meters and Base Units are cleaned & disinfected:
After each patient use (meter only) - before replacing into the meter tote box
or base unit.
When visibly soiled
Remove the meter from the base unit
Uplug Base Unit if it is to be cleaned to eliminate electrical hazard.
To Clean:
Gently wipe surfaces with Caviwipe to clean or disinfect.
Do not allow pooling on the touch screen or contact surfaces.
Same cleaning wipe can be used for all parts of meter/base unit.
meter has only one opening – strip port entry
Always allow meter and base unit to thoroughly air dry before putting back into
use
To Disinfect:
If visibly soiled wipe to remove soil then wipe with a 2
nd
wipe for disinfection.
To Disinfect / Decontaminate – allow the surface to remain wet for 3 minutes,
wipe dry.
HANDS ON - Work on a clean surface. Use paper towel or tissue. Do not place
reagents directly on surface.
(HAND OUT METERS, REAGENTS, Temporary Operator ID Bar Codes)
Performing a Quality Control Test:

Other manuals for Accu-Chek Inform II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Inform II and is the answer not in the manual?

Accu-Chek Inform II Specifications

General IconGeneral
Sample Size0.6 µL
Test Time5 seconds
Memory Capacity1000 test results
DisplayLCD
Measurement Range10 to 600 mg/dL
ConnectivityInfrared
Sample TypeCapillary whole blood
Operating Temperature16 to 32°C (61 to 90°F)
Relative Humidity10 to 90% non-condensing