User’s manual | Piezotome 2 Handpiece | J12821 | V4 |09 | 04/2018 | NO40EN050D - 9 / 12
2. 3 TROUBLESHOOTING
Refer to the table
Incorrect irrigation line connection
Check the irrigation line connection
No water at tip but
vibration present
Incorrect spray adjustment
Insufficient power -
Low vibration
Defective connector contacts of the
P
IEZOTOME
LED handpiece-cord assembly
Clean cord connector and socket unit contacts
Cut wire in handpiece cord
Return to customer service for P
IEZOTOME
LED handpiece-
cord assembly replacement
LED ring not fitted to the handpiece
Faulty LED ring connector contacts
Clean the LED ring connector contacts
Position the LED ring according to the "+" foolproofing
marks.
Faulty handpiece and/or cord connector
contacts
Clean the handpiece and/or cord connector contacts
Return to SATELEC after-sales
III - LIABILITY
The manufacturer's liability shall not be pledged in the following cases:
- Maintenance or repair by persons not authorized by the manufacturer.
- Use on an electrical installation that does not comply with the regulations in force.
- Other uses as those specified in this manual.
The manufacturer reserves the right to modify the handpiece and/or the user's manual without prior notice.