User’s manual | Piezotome 2 Handpiece | J12821 | V4 |09 | 04/2018 | NO40EN050D - 9 / 12
2. 3 TROUBLESHOOTING
Refer to the table
F
AULT
P
OSSIBLE
C
AUSE
R
EMEDY
No spray
Incorrect irrigation line connection
Check the irrigation line connection
No water at tip but
vibration present
Obstructed tip
Remove obstruction
Incorrect spray adjustment
Adjust spray
Insufficient power -
Low vibration
Worn or deformed tip
Replace tip
No ultrasonic output
Incorrect tip tightening
Tighten with wrench
Defective connector contacts of the
P
IEZOTOME
LED handpiece-cord assembly
Clean cord connector and socket unit contacts
Cut wire in handpiece cord
Return to customer service for P
IEZOTOME
LED handpiece-
cord assembly replacement
No light
LED ring not fitted to the handpiece
Fit the LED ring
Faulty LED ring
Replace the LED ring
Faulty LED ring connector contacts
Clean the LED ring connector contacts
LED ring poles reversed
Position the LED ring according to the "+" foolproofing
marks.
Faulty handpiece and/or cord connector
contacts
Clean the handpiece and/or cord connector contacts
Other
Return to SATELEC after-sales
III - LIABILITY
The manufacturer's liability shall not be pledged in the following cases:
- Maintenance or repair by persons not authorized by the manufacturer.
- Use on an electrical installation that does not comply with the regulations in force.
- Other uses as those specified in this manual.
The manufacturer reserves the right to modify the handpiece and/or the user's manual without prior notice.