    Kelly GrahamAug 3, 2025
    What to do if my Acteon Piezotome 2 LED has no ultrasonic output?
    • S
      Sara BrownAug 3, 2025
      If your Acteon Medical Equipment has no ultrasonic output, several factors could be at play. First, make sure the tip is properly tightened using a wrench. If the issue persists, check and clean the connector contacts of the PIEZOTOME LED handpiece-cord assembly. As a last resort, if there is a cut wire in the handpiece cord, the PIEZOTOME LED handpiece- cord assembly should be replaced by customer service.
  • C
    Curtis BakerAug 9, 2025
    How to fix no spray on Acteon Piezotome 2 LED?
    • B
      Brittney ReyesAug 9, 2025
      If your Acteon Medical Equipment is not spraying, the irrigation line connection might be incorrect. To solve this, check the irrigation line connection.
  • V
    Vincent RivasAug 15, 2025
    Why is there insufficient power or low vibration in my Acteon Medical Equipment?
    • K
      Kayla HernandezAug 15, 2025
      If you're experiencing insufficient power or low vibration with your Acteon Medical Equipment, it might be due to a worn or deformed tip. Replacing the tip should resolve this issue.
  • M
    martinezritaAug 21, 2025
    What to do if my Acteon Medical Equipment has no light?
    • M
      Marissa JohnsonAug 21, 2025
      If your Acteon Medical Equipment has no light, several issues could be the cause. Ensure the LED ring is correctly fitted to the handpiece, and if it is, consider replacing the LED ring if it's faulty. Also, check and clean the LED ring connector contacts, and verify that the LED ring poles are correctly positioned according to the '+' foolproofing marks. Lastly, inspect and clean the handpiece and/or cord connector contacts.
  • B
    Brandon VaughanAug 27, 2025
    What to do if Acteon Piezotome 2 LED Medical Equipment has vibration but no water at tip?
    • L
      lucasericAug 27, 2025
      If your Acteon Medical Equipment has vibration but no water at the tip, it could be due to an obstructed tip; in this case, remove the obstruction. Alternatively, the spray adjustment may be incorrect, so adjust the spray accordingly.

Summary

Installation Procedures

Handpiece-Cord Assembly Connection

Align and connect the cord connector to the unit socket until a click is heard. Refer to unit manual.

Tips Mounting

Screw and moderately tighten tips (1¼ turn) using the specific or universal wrench. Avoid overtorquing.

Maintenance and Sterilization

Handpiece-Cord Assembly Cleaning

Details on pre-disinfection/cleaning of the handpiece-cord assembly and tip, including specific steps and precautions.

Tip Cleaning Procedures

Covers chemical cleaning and mechanical cleaning processes for the tip, emphasizing immediate post-treatment and rinsing.

Drying Procedures

Dry the handpiece with a clean, non-woven single-use cloth to remove all liquid traces.

Sterilization Conditions and Process

Details on sterilization pouch requirements, autoclave type, temperature, time, pressure, and cautions for handpiece-cord assembly.

Assembly After Sterilization

Instructions for reassembling the LED ring, optical guide, and front part of the handpiece after sterilization.

Disposal and Recycling Guidelines

Procedures for proper disposal and recycling of the device and accessories, adhering to WEEE directive.

Component Replacement

Handpiece Replacement Criteria

Replace handpiece when thread fails to hold tips or casing shows damage from sterilization cycles.

Optical Guide and Light Ring Replacement

Replace optical guide and light ring if light intensity diminishes or casing is damaged by sterilization.

Troubleshooting Common Issues

Manufacturer Liability

Conditions Excluding Liability

Liability is excluded for unauthorized maintenance, non-compliant installations, or uses not specified in the manual.

Right to Modify

The manufacturer reserves the right to modify the handpiece and user's manual without prior notice.

Acteon Piezotome 2 LED Specifications

General IconGeneral
TypePiezoelectric surgical device
LED LightYes
Frequency28-36 kHz
SterilizationAutoclavable
IrrigationIntegrated irrigation system
Power Supply100-240 V, 50/60 Hz