6 Disinfection and sterilising
The instructions relating to cleaning, disinfection and sterilisation protocols for accessories supplied by SATELEC, a
company of Acteon group have been approved for each medical device and accessory. The applicable guides are listed
in the chapter Associated documentation page 3.
They can be downloaded at the following address: www.satelec.com/documents.
In all cases, the local regulations in force relating to the cleaning, disinfection and sterilisation protocols for accessories
take precedence over the information provided by SATELEC, a company of Acteon group.
6.1 Clean and disinfect the medical device
Do not immerse the handpiece.
The handpiece must be cleaned, disinfected and sterilised after each use.
The medical device must be in OFF or O stop position during cleaning and disinfecting procedures.
Refer to the instructions in the chapter Cleaning the irrigation system page 15.
Use alcohol disinfectant wipes.
Avoid using cleaning and disinfection products that contain flammable agents.
Otherwise, ensure that the product has completely evaporated or that there is no fuel left on the medical device and its
accessories before switching it on.
Do not use an abrasive product to clean the medical device.
Never apply sprays directly to the medical device to clean it. Always spray the product onto a wipe, then clean the
medical device.
6.2 Cleaning, disinfecting and sterilising accessories
Refer to the cleaning, disinfection and sterilisation instructions for accessories listed in the chapter Associated
documentation page 3.
User Manual | handpiece Newtron LED | J12611 | V4 | (07) | 01/2021 | NB62EN020D - Page 13/28