ENGLISH
Guide and declaration of the manufacturer - electromagnetic immunity
SOPRO device is intended to be used in the electromagnetic environment specified below.
The user should make sure it is used in this environment.
Immunity trial
Electrostatic discharges
EN 61000-4-2
Far transient bursts
EN 61000-4-4
Voltage shocks
EN 61000-4-5
Dips, brief outages and
power voltage variation
EN 61000-4-11
Magnetic field with the
network frequency
(50/60 Hz)
± 6 kV when
in contact
± 8 kV in the
air
± 2 kV for the
feed cables
± 1 kV for the
input/output
cables
Differential
mode ± 1 kV
Common
mode ± 2 kV
<5% U
T
- for
10 ms
40% U
T
- for
100 ms
70% U
T
- for
500 ms
<5% U
T
- for
5 s
3 A/m
± 6 kV
± 8 kV
± 2 kV
± 1 kV
± 1 kV
N.A.
<5% U
T
10 ms
<40% U
T
100 ms
<70% U
T
500 ms
<5% U
T
5 s
3 A/m
The floor should be wooden, concrete or tile. If
the floor is covered with a synthetic material, the
relative humidity should be at least 30%.
The main power supply quality should be one of a
traditional commercial or hospital environment.
The main power supply quality should be one of a
traditional commercial or hospital environment.
The main power supply quality should be one of a
traditional commercial or hospital environment.
If the user of SOPRO device requires it to continue
to operate during main power supply outages, it is
recommended SOPRO device is fed by an inverter
or a battery.
The magnetic field with the network frequency
should be at a characteristic level of a location in
a traditional commercial or hospital environment.
CEI 60601
Severity level
Compliance
level
Electromagnetic environment
Guide
Note: U
T
is the power voltage nominal value applied during the trial.
Sopro 617 • User manual
8