20
7. Error Messages/Troubleshooting
If an error occurs during a measurement, the measurement is dis-
continued and a corresponding error code is displayed. We recom-
mend you switch the device off and take a new measurement making
sure you allow time for relaxation (Example: err 3).
R
e
p
e
a
t
m
e
a
s
u
r
e
m
e
n
t
k
e
e
p
i
n
g
a
r
m
s
t
i
l
l
.
I
f
i
n
f
l
a
t
i
o
n
o
f
t
h
e
c
u
f
f
t
a
k
e
s
t
o
o
l
o
n
g
,
t
h
e
c
u
f
f
i
s
n
o
t
c
o
r
r
e
c
t
l
y
s
e
a
t
e
d
o
r
t
h
e
t
u
b
e
c
o
n
n
e
c
t
i
o
n
’
s
n
o
t
t
i
g
h
t
.
C
h
e
c
k
c
o
n
n
e
c
t
i
o
n
s
a
n
d
r
e
p
e
a
t
.
T
h
e
d
i
f
f
e
r
e
n
c
e
b
e
t
w
e
e
n
s
y
s
t
o
l
i
c
a
n
d
d
i
a
s
t
o
l
i
c
i
s
e
x
c
e
s
s
i
v
e
.
M
e
a
s
u
r
e
a
g
a
i
n
c
a
r
e
f
u
l
l
y
f
o
l
l
o
w
i
n
g
p
r
o
p
e
r
c
u
f
f
p
r
o
c
e
-
d
u
r
e
s
a
n
d
e
n
s
u
r
e
m
e
a
s
u
r
e
m
e
n
t
u
n
d
e
r
q
u
i
e
t
c
o
n
d
i
t
i
o
n
s
.
E
r
r
o
r
N
o
.
P
o
s
s
i
b
l
e
c
a
u
s
e
(
s
)
/
S
o
l
u
t
i
o
n
s
E
r
r
1
T
h
e
t
u
b
e
m
a
y
h
a
v
e
l
o
o
s
e
n
e
d
,
o
r
n
o
p
u
l
s
e
w
a
s
d
e
t
e
c
t
e
d
.
*
E
n
s
u
r
e
c
u
f
f
c
o
n
n
e
c
t
i
o
n
s
a
r
e
t
i
g
h
t
w
i
t
h
p
r
o
p
e
r
c
u
f
f
p
l
a
c
e
m
e
n
t
.
S
e
e
S
e
c
t
i
o
n
5
.
3
E
r
r
2
U
n
n
a
t
u
r
a
l
p
r
e
s
s
u
r
e
i
m
p
u
l
s
e
s
.
R
e
a
s
o
n
:
T
h
e
a
r
m
w
a
s
m
o
v
e
d
d
u
r
i
n
g
t
h
e
m
e
a
s
u
r
e
m
e
n
t
.
R
e
p
e
a
t
m
e
a
s
u
r
e
m
e
n
t
k
e
e
p
i
n
g
a
r
m
s
t
i
l
l
.
E
r
r
3
E
r
r
5
T
h
e
c
u
f
f
p
r
e
s
s
u
r
e
i
s
t
o
o
h
i
g
h
.
R
e
l
a
x
f
o
r
5
m
i
n
u
t
e
s
a
n
d
r
e
p
e
a
t
t
h
e
m
e
a
s
-
u
r
e
m
e
n
t
.
*
H
I
T
h
e
p
u
l
s
e
i
s
t
o
o
l
o
w
(
l
e
s
s
t
h
a
n
4
0
)
.
R
e
p
e
a
t
t
h
e
m
e
a
s
u
r
e
m
e
n
t
.
*
L
O
*If this or any other problem occurs repeatedly, please consult your
doctor.