23
10. Certifications
Device standard: This device is manufactured to meet the European and United
States standards for non-invasive blood pressure monitors:
EN1060-1 / 1995 EN1060-3 / 1997 EN1060-4 / 2004
AAMI/ANSI SP10
Electromagnetic Device fulfills the stipulations of the International standard
compatibility: IEC60601-1-2
Clinical testing: Clinical performance tests were carried out in the U.K.
Association Protocol and British HTP Protocol.
The B.H.S. (British Hypertension Society) clinical protocol was used to measure the accuracy of this product. ADC
®
blood pressure
units using the same measurement technology are graded “AA for systolic/diastolic accuracy by independent investigators using
the BHS protocol. This is the highest grading available for blood pressure monitors. Please see bhsoc.org for more information.
(Uses the same algorithm as ADC
®
B.H.S. gradedAA model number BP3BT0-A). US patents: 6,485,429 (Irregular Heartbeat
Feature) - 6,447,457 (ADC
®
Averaging Mode)
11. Technical Specifications
Weight: 471.5 g (with batteries)
Size: 160 (W) x 140 (L) x 98 (H) mm
Storage temperature: –20 to +5C (-4°F - 122°F)
Humidity: 15 to 90% relative humidity maximum
Operation temperature: 10 to 4C ( 50°F - 104°F)
Display: LCD-Display (Liquid Crystal Display)
Measuring method: Oscillometric
Pressure sensor: Capacitive
Measuring range: SYS/DIA: 30 to 280 mmHg
Pulse: 40 to 200 per minute
Cuff pressure display range: 0–299 mmHg
Memory: Automatically stores the last 99 measurements for 2 users
(total 198)
Measuring resolution: 1 mmHg
Accuracy: Pressure within ± 3 mmHg / pulse ± 5 % of the reading
Power source: a) 4 AA batteries, 1.5 V
b) AC adapter 6 V DC 600 mA
(voltage 4.5 V DC to 6 V DC)
Accessories: Wide range rigid cuff 8.7” 18.1” (22 - 46 cm)
Technical alterations reserved!
12. How to Contact Us
Toll Free Help Line: 1-800-ADC-2670
Email: info@adctoday.com
Made in China