Amplification ranges: Range Resolution
± 20 mV 625 nV
± 10 mV 313 nV
± 5 mV 156 nV
± 2 mV 62 nV
± 1 mV 31 nV
± 500 µV 15 nV
± 200 µV 6 nV
± 100 µV 3 nV
Gain accuracy: 1 %
Non-linearity: 1 %
Noise: < 1 µV
rms
(0.5 - 2 kHz)
Maximum input over-voltage: ± 5 V
Input leakage current: < 4 µA
rms
@ 120 V, 60 Hz
< 4 µA
rms
@ 240 V, 50 Hz
DC blocking: ±0.3 V
Baseline restoration: Automatic or user-controlled
Input impedance: 100 MΩ to isolated ground (~500 pF per lead) using
supplied Bio Amp subject cable and lead wires
Safety: IEC60601-1 approval
Isolation rating: 4000 V AC
rms
(1 minute)
IMRR: 130 dB
CMRR: 110 dB
Filters - High pass: Single pole 0.5 Hz and 10 Hz
- Low pass 10, 20, 50, 100, 250, 500, 1000, 2000 Hz
- Notch Mains
Audio output (26T only): Stereo ouput socket supplying an analog audio signal
from both bio amp channels. Suitable for earphones,
headphones and most externally powered speakers.
Output is 300 mV at full scale

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ADInstruments POWERLAB TEACHING SERIES and is the answer not in the manual?

ADInstruments POWERLAB TEACHING SERIES Specifications

General IconGeneral
BrandADInstruments
ModelPOWERLAB TEACHING SERIES
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals