Washdown
Waterline
Separator
Tank
Automated Collection Tank Drainage
During operation, the separator tank collects and removes
liquid waste from the vacuum suction, storing up to 8.5 gallons
(32 L) safely. Once maximum uid volume is reached, a oat
switch automatically shuts o the vacuum and drains the
contents to prevent overow. After draining completes and the
oat drops, the system automatically restores power in about
10 seconds for uninterrupted performance.
To manually activate the full drain and rinse cycle, power o the
vacuum unit briey. The rapid automated process takes under
2 minutes. ADS recommends manually initiating a complete
drainage and washdown at the end of each day by powering
the system o momentarily. This helps optimize equipment
hygiene and lifespan.
See the "Maintenance" section for step-by-step instructions.
TROUBLESHOOTING
WARNING
The vacuum system service panel is restricted to qualied technicians only. Customers should not attempt
to access the interior components or service the vacuum system. Please contact an authorized ADS
service provider for any repair needs to ensure proper handling by trained personnel.
This troubleshooting guide is intended to assist with diagnosing common vacuum system issues. Some conditions may
be resolved by the customer, while others require servicing by a qualied technician. For any problem involving internal
components or that persists after following the troubleshooting steps, please contact your authorized ADS dealer for
assistance. Our trained service technicians can quickly diagnose and resolve equipment issues to restore normal operation.
Condition Potential Cause Procedure/Remedy
Vacuum motor
runs continuously
with little to no
suction
Leak in the vacuum system
Turn on the vacuum and listen closely for leaks throughout
the system. Check vacuum gauge - continuous operation
below 10inHg indicates a leak. Contact authorized ADS dealer
for service.
Inspect tubing for pinches or kinks. Replace damaged tubing.
Turn o vacuum to initiate drain and rinse cycles. Contact
dealer if issue persists.
Pinched or kinked vacuum tubing Inspect tubing for pinches or kinks. Replace damaged tubing.
Stuck separator tank drain check
valve (open)
Turn o vacuum to initiate drain and rinse cycles. Contact dealer if
issue persists.
Vacuum motor
won't run while
master power
switch is on
Stuck separator tank oat switch
Turn o vacuum to initiate drain and rinse cycles. Contact dealer if
issue persists.
Vacuum is in power safe mode due
to power loss
(Note: power safe mode occurs
when the vacuum loses power and
cannot restart within 15 seconds.)
1. Turn o master power switch.
2. Disconnect power for 3 minutes.
3. Reconnect power.
4. Turn on master power switch.
Contact dealer if issue persists.