IONOMAT 1
Page 8 sur 14
4) Set the right button to "Iono 20" and gradually increase to read between 3 and 6 on the
display.
When using, do not exceed 6 mA read on the display; a higher current may cause
small burns that heal spontaneously. Wait for complete recovery before resuming
treatment.
You will read 00.0 if the electrodes are not in place under the arms or are not
sufficiently wet.
Wet the biocompatible protective sponges thoroughly.
Children (From 7 years
old)
Women
Men
Underarms
1 à 3 mA
4 à 6 mA
4 à 6 mA
Frequency of sessions: In attack treatment 3 to 5 sessions the first week, then 2 sessions per week for
4 weeks and 1 maintenance session per week all year round.
Warts treatment
7.1 Treatment of warts unilaterally (on one side)
1) Add one tablespoon of sodium salicylate to the water on the electrode side connected
to the red cable, the other end is immersed in ordinary water
2) Use the "Lock" position of the right button.
3) Then refer to the instructions for the treatment of sweating on your hands or feet.
7.2 Treatment of bilateral warts (both sides)