IONOMAT 1
Page 9 sur 14
1) Add one tablespoon of sodium salicylate to each tray.
2) Use the "Iono 30" position of the right button.
Then refer to the instructions for treating hand or foot perspiration.
Children (From 7 years
old)
Women
Men
Hands
2 à 8 mA
5 à 22 mA
8 à 25 mA
Feet
5 à 10 mA
8 à25 mA
12 à 25 mA
Frequency of sessions: Two sessions per week are necessary until the wart(s) disappears completely,
the skin must be clear
8 Cleaning and disinfection of the device
8.1 IONOMAT maintenance
Do not use acetone solvent
The cleaning of the device is limited to the external parts with a non-abrasive sponge
soaked in alcohol or a dry cloth
Never immerse the unit in water or clean it with ether.
Device :
- The cleaning of the device is limited to the external parts with a non-abrasive sponge
soaked in alcohol or a dry cloth.
- Never immerse the device in water or clean it with ether.
Accessories :
- Dry the electrodes.
- Dry the biocompatible grids
- Dry the bins after each use.
- Renew the electrodes every month with a scouring pad to maintain their shine.