• Always monitor the resistance to flow and excessive rain-out and
change the HME device as per manufacturer’s instructions.
• Do not use a filter or heat-moisture exchanger (HME) between the
nebulizer and patient airway.
• Condensate can collect and occlude ventilator circuits. Always position
ventilator circuits so that fluid condensate drains away from the patient.
• Always connect a bacteria filter to the expiratory inlet of the ventilator.
Otherwise the function of the expiratory channel may be degraded.
Optimum Use
For optimum use of the Aerogen Solo, ensure it is correctly orientated as
shown in Figure 8. This applies to both 30 Minute and Continuous modes.
Figure 10. Optimum Use of the Aerogen Solo