25
Troubleshooting
If these suggestions do not correct the problem, discontinue use of any
device and contact your local Aerogen sales representative.
Table 2. Aerogen Pro-X Controller Troubleshooting
If this happens: It could mean: Try this:
The 30 Min. indicator
flashes during
nebulization.
Batterypowerislow.
Recharge battery
(seeRechargingtheBattery).
Batterywillnot
recharge.
Controller is connected
to the AC/DC adapter
and the battery charging
light is illuminated green
and the 30 Min. indicator
light is flashing.
It may be time to replace
the battery.
Contact your local Aerogen
sales representative.
Batterywillnotretain
initial charge.
Rechargeable battery may
need to be replaced.
Contact your local Aerogen
sales representative.
The 30 Min. or
Continuous light
illuminates, but aerosol
is not visible.
No medication in nebulizer.
Refill medication through filler
cap in the nebulizer (see page
19).
It may be time to replace
the nebulizer.
See Warranty and Life of
Product. Refer to Aerogen
Solo parts list by visiting
www.aerogen.com.
30 Min. or Continuous
indicator does not light
whenOn/Offpower
button is pressed.
Thereisnopowertothe
system.
Verify that AC/DC adapter
is securely attached to
controller.
Rechargeable battery is
depleted.
Recharge battery (see
RechargingtheBattery).
The fault indicator light
illuminates.
The controller cable is
incorrectly connected to
the nebulizer, or electronics
are malfunctioning.
Verify that controller cable
is correctly connected to
both the nebulizer and the
controller.