19
Nebulization Modes
30 Minute Mode (Intermittent)
Warnings
• To avoid damage to the Aerogen Solo, do not use a syringe with a
needle.
• During use observe for correct functioning of the nebulizer.
• The maximum capacity of the nebulizer is 6 mL.
For intermittent doses less than or equal to 6 mL:
1. Open the plug on the nebulizer.
2. Use a pre-filled ampoule or syringe to add medication into the filler port
of the nebulizer (Figure 14).
3. Close the plug.
Figure 14. Filling the nebulizer with a pre-filled ampoule
4. To start a 30 Minute nebulization cycle, press and release the blue
On/Off power button (Figure 4). The green 30 Minute indicator light
illuminates to indicate that the 30 Minute nebulization cycle is in
progress.
5. To stop the nebulizer at any time, press the On/Off power button. The
indicator turns off to indicate that nebulization has stopped.
Note: Medication can be added to the Aerogen Solo during nebulization.
This does not interrupt nebulization or ventilation.