5
Do not modify this equipment without the authorization of the manufacturer.
To avoid damage to the nebulizer:
• Do not apply undue pressure to the domed aperture plate in the center
of the nebulizer.
• Do not push out the Aerogen Vibronic
®
aerosol generator.
• Do not use a syringe with a needle to add medication.
• Do not attempt to clean the nebulizer.
Do not use or store outside of specified environmental conditions.
Federal (US) Law restricts this device to sale by or on the order of a
physician.
Use of the Aerogen Solo and T-piece during the administration of volatile
anesthetics may result in adverse effects on the constituent plastics. Do
not use with volatile anesthetics unless known to be compatible. Aerogen
have determined that, using anesthetic ventilators, the following volatile
anesthetic agents are compatible under the stated conditions below:
Anesthetic Agent Proprietary Name
Maximum
Percentage of
Anesthetic
Maximum Duration
of Exposure
Isoflurane FORANE
®
3.5 % 12 hours
Sevoflurane SEVOFLURANE
®
8 % 12 hours
Desflurane SUPRANE
®
10 % 12 hours