Control solution tests should be performed when you:
[1] First get your meter.
[2] Suspect that your meter or test strips are not working
properly.
[3] Think your test results are not accurate.
[4] Have dropped or damaged your meter or exposed your
meter to liquids.
[5] Are advised by your healthcare professional to do so.
Important AgaMatrix Control Solution information:
[1] Use only AgaMatrix Control Solution with your system.
[2] Always replace the cap on the control solution bottle
immediately after use.
[3] Do not add water or any other liquid to control solution.
[4] Control solution tests should be performed within the
system operating temperature range of 50 °F to 104 °F (10 °C
to 40 °C).